Operating Instructions TDR sensor for continuous level measurement of bulk solids VEGAFLEX 82 4 … 20 mA/HART two-wire Rod and cable probe With SIL qualification Document ID: 44222
Quick start Quick start Mounting The quick start procedure enables a quick setup with many applications. You can find further information in the respective chapters of the operating instructions manual. 1. Distance from the metallic vessel wall > 300 mm. Distance from non-metallic vessel wall > 500 mm. The probe must not touch any installations or the vessel wall. 2. In non-metallic vessels, place a metal sheet beneath the process fitting. 1 2 Fig.
Quick start 2. In the menu item "Medium - Dielectric constant" you can define the medium type (medium). 3. Carry out the adjustment in the menu items "Min. adjustment" and "Max. adjustment". 4. A "Linearization" is recommended for all vessels in which the vessel volume does not increase linearly with the level - e.g. in a horizontal cylindrical or spherical tank. Activate the appropriate curve.
Contents Contents About this document 1.1 Function............................................................................................................................ 6 1.2 Target group...................................................................................................................... 6 1.3 Symbolism used................................................................................................................ 6 2 For your safety 2.1 Authorised personnel....................
Contents 9 Set up with other systems 9.1 DD adjustment programs................................................................................................ 58 9.2 Field Communicator 375, 475......................................................................................... 58 10 Diagnostics and service 10.1 Maintenance................................................................................................................... 59 10.2 Diagnosis memory.......................................
1 About this document 1 About this document 1.1 Function 1.2 Target group 1.3 Symbolism used This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectification. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device. This operating instructions manual is directed to trained specialist personnel.
2 For your safety 2 2.1 For your safety Authorised personnel All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator. During work on and with the device the required personal protective equipment must always be worn. 2.2 Appropriate use VEGAFLEX 82 is a sensor for continuous level measurement. You can find detailed information on the application range in chapter "Product description".
2 For your safety Electromagnetic compatibility Instruments with plastic housing as well as in four-wire or Ex-d-ia version are designed for use in an industrial environment. Nevertheless, electromagnetic interference from electrical conductors and radiated emissions must be taken into account, as is usual with a class A instrument according to EN 61326-1. If the instrument is used in a different environment, the electromagnetic compatibility to other instruments must be ensured by suitable measures. 2.
3 Product description 3 Type plate Product description 3.1 Configuration The nameplate contains the most important data for identification and use of the instrument: 1 2 3 4 5 6 7 8 9 17 16 15 14 13 12 10 11 Fig.
3 Product description • Enter the serial number manually in the app Scope of this operating instructions manual This operating instructions manual applies to the following instrument versions: Versions The instrument and the electronics version can be determined via the product code on the type label as well as on the electronics. • • • • Hardware from 1.0.0 Software from 1.0.0 DTM from version 1.67.2 Standard electronics: Type FX80H.
3 Product description Between a dielectric constant of 1.5 and 3, the function switches on, if required. As soon as the level echo can no longer be detected, probe end tracking is automatically activated. The measurement is continued with the last calculated dielectric constant. The accuracy thus depends on the stability of the dielectric constant. If you measure a medium with a dielectric constant below 1.5, probe end tracking is always active.
3 Product description VEGACONNECT The interface adapter VEGACONNECT enables the connection of communication-capable instruments to the USB interface of a PC. For parameter adjustment of these instruments, the adjustment software PACTware with VEGA-DTM is required. You can find further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628). VEGADIS 81 The VEGADIS 81 is an external display and adjustment unit for VEGA plics® sensors.
3 Product description The extensions are available in the following lengths: Rod: ø 16 mm (0.63 in) • Basic segments: 20 … 5900 mm (0.79 … 232 in) • Rod segments: 20 … 5900 mm (0.79 … 232 in) • Curved segments: 100 x 100 mm (3.94 … 3.94 in) You can find further information in the operating instructions manual "Rod extension VEGAFLEX series 80". Spacer If you mount the VEGAFLEX 82 in a bypass tube or standpipe, you have to avoid contact to the bypass tube by using a spacer at the probe end.
4 Mounting 4 Screwing in Mounting 4.1 General instructions On instruments with process fitting thread, the hexagon must be tightened with a suitable screwdriver. Wrench size see chapter "Dimensions". Warning: The housing must not be used to screw the instrument in! Applying tightening force can damage internal parts of the housing.
4 Mounting vessels, the distance to the vessel wall should be at least 500 mm (19.7 in). During operation, the probe must not touch any installations or the vessel wall. If necessary, fasten the probe end. In vessels with conical bottom it can be advantageous to mount the sensor in the center of the vessel, as measurement is then possible nearly down to the lowest point of the bottom. Keep in mind that measurement all the way down to the tip of the probe may not be possible. The exact value of the min.
4 Mounting 1 2 Fig. 6: Installation in non-metallic vessel 1 2 Flange Metal sheet Concrete vessel When installed in thick concrete ceilings, VEGAFLEX 82 should be mounted front flush to the lower edge. In concrete silos, the distance to the wall should be at least 500 mm (20 in). ø >160 mm (ø >6.3") Socket 16 If possible, avoid sockets. Mount the sensor flush with the vessel top. If this is not possible, use short sockets with small diameter.
4 Mounting Higher sockets or sockets with a bigger diameter can generally be used. They can, however, increase the upper blocking distance (dead band). Check if this is relevant for your measurement. In such cases, always carry out a false signal suppression after installation. You can find further information under "Setup procedure". d d h _ 150 < _ 100 < h DN40 ... DN150 > DN150 ... DN200 Fig. 8: Mounting socket When welding the socket, make sure that the socket is flush with the vessel top.
4 Mounting Fig. 10: Mounting of the sensor with inflowing medium Measuring range The reference plane for the measuring range of the sensors is the sealing surface of the thread or flange. Keep in mind that a min. distance must be maintained below the reference plane and possibly also at the end of the probe - measurement in these areas is not possible (dead band). The length of the cable can be used all the way to the end only when measuring conductive products.
4 Mounting 1 1 2 2 Fig. 11: Fasten the probe 1 2 Lateral installation Probe Retaining sleeve In case of difficult installation conditions in liquid applications, the probe can be also mounted laterally. For this purpose, adapt the rod with rod extensions or bow-shaped segments. Let the probe length determine automatically by the instrument to compensate the resulting running time changes. The determine probe length can deviate from the actual probe length when using bow-shaped segments.
5 Connecting to power supply 5 Safety instructions Voltage supply Connecting to power supply 5.1 Preparing the connection Always keep in mind the following safety instructions: • • Connect only in the complete absence of line voltage If overvoltage surges are expected, overvoltage arresters should be installed Power supply and current signal are carried on the same two-wire cable. The operating voltage can differ depending on the instrument version.
5 Connecting to power supply suppressed, but the protective effect against high frequency interference signals remains. Cable screening and grounding with potential differences Considerable potential differences exist inside galvanic plants as well as vessels with cathodic corrosion protection. Warning: Very large equalisation currents can flow through the cable screen when the screen is grounded on both ends.
5 Connecting to power supply Fig. 12: Connection steps 5 and 6 - Single chamber housing Fig. 13: Connection steps 5 and 6 - Double chamber housing 6. Insert the wire ends into the terminals according to the wiring plan Information: Solid cores as well as flexible cores with wire end sleeves are inserted directly into the terminal openings. In case of flexible cores without end sleeves, press the terminal from above with a small screwdriver; the terminal opening is freed.
5 Connecting to power supply 9. Tighten the compression nut of the cable entry. The seal ring must completely encircle the cable 10. Reinsert the display and adjustment module, if one was installed 11. Screw the housing cover back on The electrical connection is hence finished. 5.3 Wiring plan, single chamber housing The following illustration applies to the non-Ex, Ex-ia and Ex-d version. Electronics and connection compartment 2 3 4...20mA (+)1 2(-) 5 6 7 8 4 1 Fig.
5 Connecting to power supply Connection compartment 2 3 4...20mA (+)1 2(-) 5 6 7 8 4 1 Fig. 16: Connection compartment, double chamber housing 1 2 3 4 Voltage supply, signal output For display and adjustment module or interface adapter For external display and adjustment unit Ground terminal for connection of the cable screen Information: The use of an external display and adjustment unit and a display and adjustment module in parallel in the connection compartment is not supported. 5.
5 Connecting to power supply Connection compartment 4...20mA (+)1 2 2(-) 1 Fig. 18: Connection compartment, Ex-d double chamber housing 1 2 Voltage supply, signal output Ground terminal for connection of the cable screen DIS-ADAPT with plug connector M12 x 1 for external display and adjustment unit 1 2 3 Fig. 19: View to the electronics compartment 1 2 3 DIS-ADAPT Internal plug connection Plug connector M12 x 1 4 3 1 2 Fig.
5 Connecting to power supply Wire assignment, connection cable Contact pin Colour connection cable in the sensor Terminal, electronics module Pin 4 Black 8 5.6 Wiring plan - version IP 66/IP 68, 1 bar 1 2 Fig. 21: Wire assignment fix-connected connection cable 1 2 Supplementary electronics - Second current output brown (+) and blue (-) to power supply or to the processing system Shielding 5.
5 Connecting to power supply SIM-Card Status Test USB (+)1 2(-) 1 Fig. 23: Radio module PLICSMOBILE integrated in the connection compartment 1 Voltage supply You can find detailed information on connection in the supplementary instructions "PLICSMOBILE GSM/GPRS radio module". 5.8 Switch-on phase After connecting the instrument to power supply or after a voltage recurrence, the instrument carries out a self-check for approx.
6 Functional safety (SIL) 6 Functional safety (SIL) 6.1 Objective In case of dangerous outages, processing facilities and machines can cause risks for persons, environment and property. The risk of such outages must be judged by the plant operator. Dependent thereon are measures for risk reduction through error prevention, error detection and fault control. The part of plant security depending on the correct functioning of safety-related components for risk reduction is called functional safety.
6 Functional safety (SIL) • Tool for operation and parameterization 4 … 20 mA current output 6.4 Safety concept of the parameterization The following tools are permitted for parameterization of the safety function: • • The integrated display and adjustment unit for on-site adjustment The DTM suitable for the signal conditioning instrument in conjunction with an adjustment software according to the FDT/DTM standard, e. g. PACTware Note: For operation of the VEGAFLEX 82 the DTM Collection version 1.
6 Functional safety (SIL) Change parameter All parameters changed by the operator are automatically stored temporarily so that they can be verified in the next step. Verify parameters/Lock adjustment After setup, the modified parameters must be verified (confirm the correctness of the parameters). To do this, you first have to enter the PIN. Here the adjustment is locked automatically. Then you carry out a comparison of two character strings. You must confirm that the character strings are identical.
6 Functional safety (SIL) Setup is then finished. Function test required Should a function test be necessary, the following message is displayed on the display and adjustment module. The adjustment software also signals that a function test is required. If a function test is required, the switching point or the range must be controlled with the original medium. For this purpose, you have to decide for your application which condition is potentially critical.
6 Functional safety (SIL) Measure the current output and compare the value with the calculated current value. You have to determine the permissible deviation of the values yourself. This deviation depends on the the accuracy requirements of your measurement loop. Determine the permissible tolerance for the deviation. Mode "Range monitoring" If both levels are important for the safety function, you have to proceed according to the mode "Range monitoring". 1.
7 Set up with the display and adjustment module 7 7.1 Set up with the display and adjustment module Insert display and adjustment module The display and adjustment module can be inserted into the sensor and removed any time. Four positions displaced by 90° can be selected. It is not necessary to interrupt the power supply. Proceed as follows: 1. Unscrew the housing cover 2. Place the display and adjustment module in the requested position onto the electronics and turn to the right until it snaps in 3.
7 Set up with the display and adjustment module 1 2 Fig. 25: Insertion of the display and adjustment module into the double chamber housing 1 2 In the electronics compartment In the connection compartment (with Ex-d-ia version not possible) Note: If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication, a higher cover with an inspection glass is required. 7.
7 Set up with the display and adjustment module • • • –– Confirm selected menu –– Edit parameter –– Save value [->] key: –– Presentation, change measured value –– Select list entry –– Select editing position [+] key: –– Change value of the parameter [ESC] key: –– Interrupt input –– Jump to next higher menu Adjustment system The sensor is adjusted via the four keys of the display and adjustment module. The LC display indicates the individual menu items.
7 Set up with the display and adjustment module Setup: Settings, e.g. measurement loop name, medium, vessel, adjustment, signal output, device unit, false signal suppression, linearization curve Display: Settings, e.g., for language, measured value display, lighting Diagnosis: Information, e.g.
7 Set up with the display and adjustment module correctly. If necessary, other measures must be taken to maintain the safety function. Setup - Measurement loop name Here you can assign a suitable measurement loop name. Push the "OK" key to start the processing. With the "+" key you change the sign and with the "->" key you jump to the next position. You can enter names with max. 19 characters.
7 Set up with the display and adjustment module You have the option of choosing the demonstration mode. In this mode, the sensor ignores the parameters of the application and reacts immediately to any change. This mode is only suitable for test and demonstration purposes and must not be used in a safety-instrumented application (SIL). Setup - Medium, dielectric constant In this menu item, you can define the medium type (product).
7 Set up with the display and adjustment module Adjust the requested percentage value with [+] and store with [OK]. Enter the suitable distance value in m for the empty vessel (e.g. distance from the flange to the probe end) corresponding to the percentage value. The distance refers tot he sensor reference plane (seal surface of the process fitting). Setup - Damping To damp process-dependent measured value fluctuations, set an integration time of 0 … 999 s in this menu item.
7 Set up with the display and adjustment module D -h +h For the socket correction you have to enter the height of the socket above the upper edge of the vessel. If the socket is lower than the upper edge of the vessel, this value can also be negative. Fig. 27: Vessel height und socket correction value D Vessel height +h Positive socket correction value -h Negative socket correction value Since scaling is very extensive, scaling of the level value was divided into two menu items.
7 Set up with the display and adjustment module Setup - Scaling Level 2 Since scaling is very extensive, scaling of the level value was divided into two menu items. In menu item "Level 2" you define the scaling format on the display and the scaling of the measured level value for 0 % and 100 %. Setup - Current output Size In menu item"Current output, size" you determine which measured value the current output refers to.
7 Set up with the display and adjustment module The default setting is min. current 3.8 mA and max. current 20.5 mA. Setup - Current output 2 If a supplementary electronics with a second current output is implemented in the instrument, then you can adjust the second current output separately. In menu item"Current output 2" you specify which measured value the second current output refers to. The second current output cannot be used as an output in the sense of a safety-instrumented application (SIL).
7 Set up with the display and adjustment module Note: A false signal suppression is only recommended with liquid applications. A false signal suppression detects, marks and saves these false signals so that they are no longer taken into account for level measurement. This should be done with the lowest possible level so that all potential interfering reflections can be detected. Proceed as follows: Enter the actual distance from the sensor to the product surface.
7 Set up with the display and adjustment module For this reason, the instrument is shipped in locked conditon. The PIN in the delivery status is "0000". Character string comparison and serial number You first have to carry out the character string comparison. This is used to check the character respresentation. Confirm if the two character strings are identical. The verification texts are provided in German and with all other menu languages in English.
7 Set up with the display and adjustment module Verify parameter All safety-relevant parameters must be verified after a change. After the function test, all modified, safety-relevant parameters will be listed. Confirm the modified values one after the other. If the described process of parameter adjustment was run through completely and correctly, the instrument will be locked and hence ready for operation. Otherwise the instrument remains in the released and hence unsafe condition.
7 Set up with the display and adjustment module Display - Displayed value 1 In this menu item, you define the indication of the measured value on the display. You can display two different measured values. In this menu item, you define measured value 1. The default setting for the displayed value 1 is "Filling height Level". Display - Displayed value 2 In this menu item, you define the indication of the measured value on the display. You can display two different measured values.
7 Set up with the display and adjustment module Diagnostics - Peak values The respective min. and max. measured values are saved in the Measurement certainty sensor. The two values are displayed in the menu item "Peak values, measurement certainty". The measurement can be influenced by the process conditions. In this menu item, the measurement certainty of the level measurement is displayed as percentage value. The higher the value, the more reliable the measurement.
7 Set up with the display and adjustment module Diagnosis - Simulation In this menu item you can simulate measured values via the current output. This allows the signal path to be tested, e.g. via downstream indicating instruments or the input card of the control system. Select the requested simulation size and adjust the requested value. Caution: During simulation, the simulated value is outputted as 4 … 20 mA current value and digital HART signal. Push the [ESC] key to deactivate the simulation.
7 Set up with the display and adjustment module Diagnosis - Proof test With the function "Proof test", the function of the instrument can be checked on a recurring basis. During the function test, the safety function must be treated as unsafe. Keep in mind that the function test influences downstream connected devices. You can find detailed information on the proof test in the Safety Manual (SIL).
7 Set up with the display and adjustment module signal suppression, user-programmable linearization curve as well as the measured value memory will be deleted. Basic settings: Resetting of the parameter settings incl. special parameters to the default values (presettings) of the respective instrument. Any created false signal suppression or user-programmable linearization curve as well as the measured value memory are deleted. The following tables show the default values of the instrument.
7 Set up with the display and adjustment module Menu item Default value Linearization - Vessel height Probe length Scaling size - Level Volume in l Scaling unit - Level Litres Scaling format - Level Without decimal positions Scaling level - 100 % correspond to 100 Scaling level - 0 % correspond to 0 Accept scaling of the level measurement Yes Scaling variable - Interface Volume Scaling unit - Interface Litres Scaling format - Interface Without decimal positions Scaling interface - 100
7 Set up with the display and adjustment module Menu item Default value Status signals - Out of specification Switched off Status signals - Maintenance Switched on Device memory - Echo curve memory Stopped Device memory - Measured value memory Started Device memory - Measured value memory Measured values Distance level, percentage value level, reliability level, electronics temperature Device memory - Measured value memory - Recording in time interval 3 min.
7 Set up with the display and adjustment module power failure. From there, they can be written into one or more sensors or kept as backup for a possible electronics exchange. Note: Before the data are stored in the sensor, a check is carried out to determine if the data fit the sensor. If the data do not fit, a fault signal is triggered or the function is blocked. When data are being written into the sensor, the display shows which instrument type the data originate from and which TAG-no. this sensor had.
7 Set up with the display and adjustment module Info - Factory calibration date In this menu item, the date of factory calibration of the sensor as well as the date of the last change of sensor parameters are displayed via the display and adjustment module or via the PC. Info - Sensor characteristics In this menu item, the features of the sensor such as approval, process fitting, seal, measuring range, electronics, housing and others are displayed. 7.
8 Setup with PACTware 8 Via the interface adapter directly on the sensor Setup with PACTware 8.1 Connect the PC 2 1 3 Fig. 28: Connection of the PC directly to the sensor via the interface adapter 1 2 3 USB cable to the PC Interface adapter VEGACONNECT Sensor Via the interface adapter and HART 2 4 USB N OPE 3 TWIST LO CK 1 5 Fig.
8 Setup with PACTware such cases, the interface converter can be connected parallel to the 4 … 20 mA cable (dashed line in the previous illustration). Prerequisites 8.2 Parameter adjustment with PACTware For parameter adjustment of the sensor via a Windows PC, the configuration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required. The up-to-date PACTware version as well as all available DTMs are compiled in a DTM Collection.
8 Setup with PACTware In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves. The standard version is available as a download under www.vega.com/downloads and "Software". The full version is available on CD from the agency serving you. 8.
9 Set up with other systems 9 9.1 Set up with other systems DD adjustment programs Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™ and PDM. The files can be downloaded at www.vega.com/downloads under "Software". 9.2 Field Communicator 375, 475 Device descriptions for the instrument are available as EDD for parameter adjustment with the Field Communicator 375 or 475.
10 Diagnostics and service 10 Diagnostics and service 10.1 Maintenance If the device is used correctly, no maintenance is required in normal operation. When used in safety-instrumented systems (SIS), the safety function must be carried out on the instrument in regular time intervals by means of a proof test. Hence possible undetected, dangerous failure can be identified. The operator's responsibility to select the kind of test. The time intervals depend on the used PFDAVG.
10 Diagnostics and service • • • Switch-on and switch-off times Status messages (according to NE 107) Error messages (according to NE 107) The data are read out via a PC with PACTware/DTM or the control system with EDD. Echo curve memory The echo curves are stored with date and time and the corresponding echo data. The memory is divided into two sections: Echo curve of the setup: This is used as reference echo curve for the measurement conditions during setup.
10 Diagnostics and service Failure: Due to a malfunction in the instrument, a failure message is outputted. This status message is always active. It cannot be deactivated by the user. Function check: The instrument is in operation, the measured value is temporarily invalid (for example during simulation). This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
10 Diagnostics and service Code Cause Rectification F080 –– General software error –– Separate operating voltage briefly F105 –– The instrument is still in the start phase, the measured value could not yet be determined –– Wait for the end of the switch-on phase –– Duration depending on the version and parameter adjustment max. 5 min.
10 Diagnostics and service Code Cause Rectification C700 –– A simulation is active –– Finish simulation –– Wait for the automatic end after 60 mins. C701 –– Parameter verification was interrupted –– Finish parameter verification Text message Simulation active Parameter verification Out of specification The following table shows the error codes and text messages in the status message "Out of specification" and provides information on causes as well as corrective measures.
10 Diagnostics and service Code Cause Rectification M503 Reliability too low –– The echot/noise ratio is the small for a reliable measurement –– Check installation and process conditions –– Clean the antenna –– Change polarisation direction –– Use instrument with higher sensitivity M504 –– Hardware defect –– Check connections –– Exchanging the electronics –– Send instrument for repair M505 –– Level echo can no longer be detected –– Clean the antenna –– Use a more suitable antenna/sensor –– Remov
10 Diagnostics and service Error Cause –– Check connection according to chapter "Connection steps" and if necessary, correct according to chapter "Wiring plan" –– Voltage supply missing –– Check cables for breaks; repair if necessary –– Operating voltage too low or load resistance too high –– Check, adapt if necessary –– Electronics module in the sensor defective –– Exchange the instrument or send it in for repair Current signal greater than 22 mA or less than 3.
10 Diagnostics and service Measurement error with constant level 1. Measured value shows a too low or too high level Error pattern Level Fault description 0 Level 2. Measured value jumps towards 100 % time 0 time Cause Rectification –– Min./max. adjustment not correct –– Adapt min./max.
10 Diagnostics and service Measurement error during emptying 8. Measured value remains reproducible in one position during emptying Error pattern 0 Cause time –– Stored false signals in this position are larger than the level echo 0 Rectification –– False echo larger than the level –– Eliminate false signals in the echo close range –– Level echo too small –– Remove contamination on the probe. After having removed the source of the false signals, the false signal suppression must be deleted.
10 Diagnostics and service After an electronics exchange the settings of the electronics must be checked and verified. Only then, the instrument is again ready for operation. 10.6 Exchange or shorten cable/rod Exchanging the cable/rod The cable or rod (meas. part) of the probe can be shortened, if necessary. To loosen the rod or cable you need a fork spanner with spanner width 13. 1.
10 Diagnostics and service Information: Please keep the stated torque so that the max. tensile strength remains. 6. Enter new probe length and if necessary the new probe type and then carry out a fresh adjustment (see "Setup procedure, Carrying out min. adjustment - Carrying out max. adjustment"). Shorten cable/rod The rod or cable of the probe can be shortened individually. 1. Mark the requested length with mounted measuring rod. 2.
10 Diagnostics and service 3 150 mm (5.91") 150 mm (5.91") 70 mm (2.76") 45 mm (1.77") 1 B A 2 2 Fig. 42: Shortening the cable probe A B 1 2 3 Gravity weight - cable ø 4 mm Gravity weight - cable ø 6 mm Pins Thread M12 for eye-bolt Pins 10.
10 Diagnostics and service 10.8 How to proceed in case of repair You can find a repair form as well as detailed information on how to proceed under www.vega.com/downloads and "Forms and certificates". By doing this you help us carry out the repair quickly and without having to call back for needed information.
11 Dismounting 11 Dismounting 11.1 Dismounting steps Warning: Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel or pipeline, high temperatures, corrosive or toxic products etc. Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order. 11.2 Disposal The instrument consists of materials which can be recycled by specialised recycling companies.
12 Supplement 12 Supplement 12.1 Technical data General data 316L corresponds to 1.4404 or 1.4435 Materials, wetted parts ƲƲ Process fitting ƲƲ Process seal on the instrument side (cable/rod leadthrough) ƲƲ Process seal ƲƲ Inner conductor (up to the separation cable/rod) 316L and PPS GF 40, Hastelloy C22 (2.4602) and PPS GF 40 FKM (SHS FPM 70C3 GLT), FFKM (Kalrez 6375), EPDM (A+P 75.5/KW75F) On site (instruments with thread: Klingersil C-4400 is attached) 316L ƲƲ Rod: ø 16 mm (0.
12 Supplement ƲƲ Instrument weight (depending on process fitting) approx. 0.8 … 8 kg (0.176 … 17.64 lbs) ƲƲ Cable: ø 4 mm (0.157 in) approx. 78 g/m (0.84 oz/ft) ƲƲ Rod: ø 16 mm (0.63 in) approx. 1580 g/m (17 oz/ft) ƲƲ Cable: ø 6 mm (0.236 in) - PA coated approx. 180 g/m (1.9 oz/ft) ƲƲ Cable: ø 6 mm (0.236 in) approx. 80 g/m (0.86 oz/ft) ƲƲ Cable: ø 11 mm (0.433 in) - PA coated approx. 320 g/m (3.44 oz/ft) ƲƲ Gravity weight for cable ø 4 mm (0.157 in) and cable: ø 6 mm (0.
12 Supplement 1 1 12 3 10 2 0 6 1 2 A 5 3 6 4 4 4 5 6 3 5 4 8 6 0 5 10 15 20 25 30 2 35 00 5 10 15 20 25 30 2 35 B Fig. 43: Max. tensile load with cereals and plastic granules - Cable: ø 4 mm (0.157 in) A B 1 2 3 4 5 6 Cereals Plastic granules Tensile force in kN (the determined value must be multiplied with safety factor 2) Cable length in m Vessel diameter 12 m (39.37 ft) Vessel diameter 9 m (29.53 ft) Vessel diameter 6 m (19.69 ft) Vessel diameter 3 m (9.
12 Supplement 1 1 18 16 8 4 2 0 5 0 5 10 15 20 25 30 35 5 4 6 6 4 6 5 10 A 7 4 12 3 8 3 14 9 3 6 2 1 2 0 0 5 10 15 20 25 30 35 2 B Fig. 45: Max. tensile load with cereals and plastic granules - Cable: ø 6 mm/ø 11 mm - PA coated A B 1 2 3 4 5 6 Cereals Plastic granules Tensile force in kN (the determined value must be multiplied with safety factor 2) Cable length in m Vessel diameter 12 m (39.37 ft) Vessel diameter 9 m (29.53 ft) Vessel diameter 6 m (19.
12 Supplement ƲƲ Cable: ø 6 mm (0.236 in) - PA coated ƲƲ Cable: ø 6 mm (0.236 in) 8 Nm (5.9 lbf ft) 20 Nm (14.75 lbf ft) ƲƲ Cable: ø 11 mm (0.433 in) - PA coated 20 Nm (14.75 lbf ft) ƲƲ Rod: ø 16 mm (0.63 in) 20 Nm (14.75 lbf ft) ƲƲ Plastic housing max. 10 Nm (7.376 lbf ft) Torque for NPT cable glands and Conduit tubes ƲƲ Aluminium/Stainless steel housing Input variable Measured variable Level of solids εr ≥ 1.5 44222-EN-130823 Min. dielectric constant of the medium max. 50 Nm (36.
12 Supplement 1 1 4 4 2 2 3 3 5 5 Fig. 47: Measuring ranges - VEGAFLEX 82 1 2 3 4 5 Reference plane Probe length L Measuring range Upper dead band (see diagrams under Accuracy - grey section) Lower dead band (see diagrams under Accuracy - grey section) Output variable Output signal 4 … 20 mA/HART Fulfilled HART specification 7 Range of the output signal 0.3 µA Failure signal current output (adjustable) ≥ 21 mA, ≤ 3.6 mA Max. output current Starting current 78 21.
12 Supplement Load Damping (63 % of the input variable), adjustable see load diagram under Power supply 0 … 999 s HART output values according to HART 7 (default setting)2) ƲƲ First HART value (PV) Linearised percentage value, level ƲƲ Third HART value (TV) Measurement certainty, level ƲƲ Second HART value (SV) ƲƲ Fourth HART value (QV) Distance to the level Electronics temperature Indication value - DIsplay and adjustment module3) ƲƲ Displayed value 1 ƲƲ Displayed value 2 Resolution, digital F
12 Supplement ƲƲ Medium Bulk solids - cereals, flour, cement (dielectric constant ~2.0) ƲƲ Installation Sensor parameter adjustment Typical deviation5) Probe end does not touch the vessel bottom No gating out of false signals carried out See following diagrams L 15 mm (0.591") 2 mm (0.079") 0 -2 mm (-0.079") -10 mm (-0.394") 1 0,08 m (3.15") 0,3 m (11.811") 0,02 m (0.787") Fig. 48: Deviation VEGAFLEX 82 in rod version 1 L Dead band - no measurement possible in this area Probe length L 15 mm (0.
12 Supplement Temperature drift - Digital output ±3 mm/10 K relating to the max. measuring range or max. 10 mm (0.394 in) Additional deviation through electromag- < ±10 mm (< ±0.394 in) netic interference acc. to EN 61326 Specifications apply also to the current output6) Temperature drift - Current output Deviation on the current output through analogue/digital conversion ±0.03 %/10 K relating to the 16 mA span max. ±0.
12 Supplement Process conditions For the process conditions, please also note the specifications on the type label. The lowest value always applies. The measurement error from the process conditions is in the specified pressure and temperature range of below 1 %. Process pressure Vessel pressure relating to the flange nominal stage -1 … +40 bar/-100 … +4000 kPa (-14.
12 Supplement 1 80°C (176°F) 3 4 55°C (131°F) -20°C (-68°F) 2 0°C (32°F) 150°C (302°F) 200°C (392°F) -40°C (-104°F) Fig. 51: Ambient temperature - process temperature, version with temperature adapter Cable version with ø 4 mm (0.157 in) Cable version, PA coated with ø 6 mm (0.
12 Supplement 1 80°C (176°F) 3 4 65°C (149°F) 2 0°C (32°F) -20°C (-68°F) 170°C (338°F) 200°C (392°F) -40°C (-104°F) Fig. 53: Ambient temperature - process temperature, version with temperature adapter Cable version with ø 6 mm (0.236 in) Cable version, PA coated with ø 11 mm (0.
12 Supplement ƲƲ Massive wire, cord ƲƲ Stranded wire with end sleeve 0.2 … 2.5 mm² (AWG 24 … 14) 0.2 … 1.5 mm² (AWG 24 … 16) Electromechanical data - version IP 66/IP 68 (1 bar) Options of the cable entry ƲƲ Cable gland ƲƲ Cable diameter (options) ƲƲ Cable entry ƲƲ Blind plug Connection cable 5 … 9 mm (0.20 … 0.35 in) 6 … 12 mm (0.24 … 0.47 in) 10 … 14 mm (0.39 … 0.55 in) ½ NPT M20 x 1.5; ½ NPT ƲƲ Wire cross-section 0.
12 Supplement Accuracy Voltage supply Operating voltage ±1 °C (1.8 °F) ƲƲ Non-Ex instrument, Ex-d instrument 9.6 … 35 V DC ƲƲ Ex-d-ia instrument 15 … 35 V DC ƲƲ Ex-ia instrument 9.6 … 30 V DC Operating voltage with illuminated display and adjustment module ƲƲ Non-Ex instrument, Ex-d instrument 16 … 35 V DC ƲƲ Ex-d-ia instrument 20 … 35 V DC ƲƲ Ex-ia instrument Interpolation protection 16 … 30 V DC Integrated Permissible residual ripple - Non-Ex, Ex-ia instrument ƲƲ for 9.
12 Supplement Plastic housing ~ 69 mm (2.72") ø 79 mm (3.11") ~ 84 mm (3.31") ø 79 mm (3.11") M20x1,5/ ½ NPT 112 mm (4.41") 112 mm (4.41") M16x1,5 M20x1,5/ ½ NPT 1 2 Fig. 54: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher 1 2 Single chamber version Double chamber version Aluminium housing ~ 116 mm (4.57") ~ 87 mm (3.43") ø 86 mm (3.39") ø 86 mm (3.39") 120 mm (4.72") 116 mm (4.
12 Supplement Aluminium housing in protection rating IP 66/IP 68 (1 bar) ~ 105 mm (4.13") ~ 150 mm (5.91") ø 84 mm (3.31") ø 84 mm (3.31") 120 mm (4.72") 116 mm (4.57") M16x1,5 M20x1,5 M20x1,5 M20x1,5/ ½ NPT 1 2 Fig. 56: Housing versions in protection IP␣66/IP␣68 (1␣bar) - with integrated display and adjustment module the housing is 9␣mm/0.35␣in higher 1 2 Single chamber version Double chamber version Stainless steel housing ~ 87 mm (3.43") ~ 69 mm (2.72") ø 79 mm (3.11") ~ 59 mm (2.
12 Supplement Stainless steel housing in protection rating IP 66/IP 68, 1 bar ~ 93 mm (3.66") ~ 103 mm (4.06") ø 80 mm (3.15") ~ 105 mm (4.13") ø 79 mm (3.11") ø 86 mm (3.39") M20x1,5/ ½ NPT 120 mm (4.72") 112 mm (4.41") 117 mm (4.61") M16x1,5 M20x1,5 1 2 M20x1,5/ ½ NPT 3 Fig. 58: Housing versions in protection IP␣66/IP␣68 (1␣bar) - with integrated display and adjustment module the housing is 9␣mm/0.
12 Supplement 46 mm (1.81") VEGAFLEX 82, cable version ø 4 mm (0.157 in)/ø 6 mm (0.236 in) - PA coated SW 36 mm (1.42") 22 mm (0.87") G¾ 3 ø 6 mm (0.24") 150 mm (5.91") L ø 4 mm (0.16") ø 20 mm (0.79") 1 2 Fig. 59: VEGAFLEX 82, cable ø 4 mm (0.157 in)/ø 6 mm (0.236 in) threaded version with gravity weight (all gravity weights with thread M12 for eye-bolt) L 1 2 3 Sensor length, see chapter "Technical data" Cable ø 4 mm (0.157 in) Cable ø 6 mm (0.
12 Supplement 80 mm (3.15") VEGAFLEX 82, cable version ø 6 mm (0.236 in)/ø 11 mm (0.433 in) - PA coated SW 55 mm (2.17") 30 mm (1.18") G1½ ø 30 mm (1.18") ø 11 mm (0.43") 150 mm (5.91") ø 6 mm (0.24") L 3 1 2 44222-EN-130823 Fig. 60: VEGAFLEX 82, cable ø 6 mm (0.236 in)/ø 11 mm (0.433 in) threaded version with gravity weight (all gravity weights with thread M12 for eye-bolt) L 1 2 3 Sensor length, see chapter "Technical data" Cable ø 6 mm (0.236 in) Cable ø 11 mm (0.
12 Supplement 80 mm (3.15") VEGAFLEX 82, rod version ø 16 mm (0.63 in) SW 55 mm (2.17") 30 mm (1.18") G1½ L 1 ø 16 mm (0.63") Fig. 61: VEGAFLEX 82, rod ø 16 mm (0.
12 Supplement 12.3 Industrial property rights VEGA product lines are global protected by industrial property rights. Further information see www.vega.com. Only in U.S.A.: Further information see patent label at the sensor housing. VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte. Nähere Informationen unter www.vega.com. Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle. Pour plus d'informations, on pourra se référer au site www.
INDEX INDEX A Accessories ––Display and adjustment module 11 ––External display and adjustment unit 12 ––Flanges 12 ––Interface adapter 12 ––Protective cap 12 Adjustment ––Max. adjustment 38 ––Min. adjustment 38 Adjustment system 35 Application 37, 38 Application area 10 B Backlight 46 Block adjustment 36 C Cable entry 20 Check signal 64 Connection cable 20 Connection procedure 21 Connection technology 21 Copy sensor settings 52 Current output 2 42 Current output Min./Max.
INDEX Probe type 53 Proof test 49 R Read out info 53 Release adjustment 43 Repair 71 Replacement parts ––Display and adjustment module with heating 12 ––Electronics module 12 ––Rod extension 12 ––Spacer 13 Reset 49 S Scaling measured value 40, 41 Sensor characteristics 54 Sensor status 46 Service hotline 67 Shielding 20 SIL 28 Simulation 48 Special parameters 53 Status messages 60 Storage 11 T Type plate 9 U Units 37 V 44222-EN-130823 Verify parameter 32 Voltage supply 20 VEGAFLEX 82 • 4 … 20 m
All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice © VEGA Grieshaber KG, Schiltach/Germany 2013 VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.