Operating Instructions TDR sensor for continuous level and interface measurement of liquids VEGAFLEX 83 4 … 20 mA/HART two-wire Rod probe - Polished version With SIL qualification Document ID: 44225
Quick start Quick start Mounting The quick start procedure enables a quick setup with many applications. You can find further information in the respective chapters of the operating instructions manual. 1. Distance from the metallic vessel wall > 300 mm. Distance from non-metallic vessel wall > 500 mm. The probe must not touch any installations or the vessel wall. 2. In non-metallic vessels, place a metal sheet beneath the process fitting. 1 2 Fig.
Quick start 2. In the menu item "Medium - Dielectric constant" you can define the medium type (medium). 3. Carry out the adjustment in the menu items "Min. adjustment" and "Max. adjustment". 4. A "Linearization" is recommended for all vessels in which the vessel volume does not increase linearly with the level - e.g. in a horizontal cylindrical or spherical tank. Activate the appropriate curve. 5.
Quick start 2. Select the parameter of the current output and the output characteristics in the menu item "Current output".
Contents 44225-EN-130823 Contents 1 About this document 1.1 Function............................................................................................................................ 7 1.2 Target group...................................................................................................................... 7 1.3 Symbolism used................................................................................................................ 7 2 For your safety 2.
Contents 9 Set up with other systems 9.1 DD adjustment programs................................................................................................ 62 9.2 Field Communicator 375, 475......................................................................................... 62 10 Diagnostics and service 10.1 Maintenance................................................................................................................... 63 10.2 Diagnosis memory.......................................
1 About this document 1 About this document 1.1 Function 1.2 Target group 1.3 Symbolism used This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectification. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device. This operating instructions manual is directed to trained specialist personnel.
2 For your safety 2 2.1 For your safety Authorised personnel All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator. During work on and with the device the required personal protective equipment must always be worn. 2.2 Appropriate use VEGAFLEX 83 is a sensor for continuous level measurement. You can find detailed information on the application range in chapter "Product description".
2 For your safety Electromagnetic compatibility Instruments with plastic housing as well as in four-wire or Ex-d-ia version are designed for use in an industrial environment. Nevertheless, electromagnetic interference from electrical conductors and radiated emissions must be taken into account, as is usual with a class A instrument according to EN 61326-1. If the instrument is used in a different environment, the electromagnetic compatibility to other instruments must be ensured by suitable measures. 2.
3 Product description 3 Type plate Product description 3.1 Configuration The nameplate contains the most important data for identification and use of the instrument: 1 17 16 2 15 3 14 4 5 6 7 8 9 13 12 11 10 Fig.
3 Product description • Enter the serial number manually in the app Scope of this operating instructions manual This operating instructions manual applies to the following instrument versions: Versions The instrument and the electronics version can be determined via the product code on the type label as well as on the electronics. • • • • Hardware from 1.0.0 Software from 1.0.0 DTM from version 1.67.2 Standard electronics: Type FX80H.
3 Product description 1 d h Fig. 5: Level measurement 1 d h Sensor reference plane (seal surface of the process fitting) Distance to the interface (HART value 1) Height - Level Probe end tracking To increase sensitivity, the probe is equipped with probe end tracking. In products with a low dielectric constant, this function is very helpful. This is the case, for example, in plastic granules, packing chips or in vessels with fluidized products. Between a dielectric constant of 1.
3 Product description h2 h1 L3 d2 L2 TS 1 d1 L1 Fig.
3 Product description Output signal The instrument is always preset to the application "Level measurement". For the interface measurement, you can select the requested output signal with the setup. Packaging 3.3 Packaging, transport and storage Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test based on ISO 4180. The packaging of standard instruments consists of environmentfriendly, recyclable cardboard.
3 Product description You can find further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628). VEGADIS 81 The VEGADIS 81 is an external display and adjustment unit for VEGA plics® sensors. For sensors with double chamber housing the interface adapter "DISADAPT" is also required for VEGADIS 81. You can find further information in the operating instructions "VEGADIS 81" (Document-ID 43814).
4 Mounting 4 Protection against moisture Mounting 4.1 General instructions Protect your instrument further through the following measures against moisture penetration: • • • Use the recommended cable (see chapter "Connecting to power supply") Tighten the cable gland Loop the connection cable downward in front of the cable gland This applies particularly to: • • • Protective caps Outdoor mounting Installations in areas where high humidity is expected (e.g.
4 Mounting measurement all the way down to the tip of the probe may not be possible. The exact value of the min. distance (lower dead band) is stated in chapter "Technical data". Fig. 7: Vessel with conical bottom Type of vessel Plastic vessel/Glass vessel The guided microwave principle requires a metal surface on the process fitting. Therefore use in plastic vessels etc.
4 Mounting Socket If possible, avoid sockets. Mount the sensor flush with the vessel top. If this is not possible, use short sockets with small diameter. Higher sockets or sockets with a bigger diameter can generally be used. They can, however, increase the upper blocking distance (dead band). Check if this is relevant for your measurement. In such cases, always carry out a false signal suppression after installation. You can find further information under "Setup procedure". d DN40 ... DN150 > DN150 ..
4 Mounting Inflowing medium Do not mount the instruments in or above the filling stream. Make sure that you detect the product surface, not the inflowing product. Fig. 11: Mounting of the sensor with inflowing medium Measuring range The reference plane for the measuring range of the sensors is the sealing surface of the thread or flange. Keep in mind that a min.
4 Mounting Rod extension In case of difficult installation conditions, for example in a socket, the probe can be adapted respectively with a rod extension. Let the probe length determine automatically by the instrument to compensate the resulting running time changes. You can find further information in the supplementary instructions of the rod extension. Autoclaved version For use in an autoclave, e.g. for sterilization, the VEGAFLEX 83 is available as autoclaved version.
5 Connecting to power supply 5 Safety instructions Voltage supply Connecting to power supply 5.1 Preparing the connection Always keep in mind the following safety instructions: • • Connect only in the complete absence of line voltage If overvoltage surges are expected, overvoltage arresters should be installed Power supply and current signal are carried on the same two-wire cable. The operating voltage can differ depending on the instrument version.
5 Connecting to power supply suppressed, but the protective effect against high frequency interference signals remains. Cable screening and grounding with potential differences Considerable potential differences exist inside galvanic plants as well as vessels with cathodic corrosion protection. Warning: Very large equalisation currents can flow through the cable screen when the screen is grounded on both ends.
5 Connecting to power supply Fig. 13: Connection steps 5 and 6 - Single chamber housing Fig. 14: Connection steps 5 and 6 - Double chamber housing 6. Insert the wire ends into the terminals according to the wiring plan 44225-EN-130823 Information: Solid cores as well as flexible cores with wire end sleeves are inserted directly into the terminal openings. In case of flexible cores without end sleeves, press the terminal from above with a small screwdriver; the terminal opening is freed.
5 Connecting to power supply 9. Tighten the compression nut of the cable entry. The seal ring must completely encircle the cable 10. Reinsert the display and adjustment module, if one was installed 11. Screw the housing cover back on The electrical connection is hence finished. 5.3 Wiring plan, single chamber housing The following illustration applies to the non-Ex, Ex-ia and Ex-d version. Electronics and connection compartment 2 3 4...20mA (+)1 2(-) 5 6 7 8 4 1 Fig.
5 Connecting to power supply Connection compartment 2 3 4...20mA (+)1 2(-) 5 6 7 8 4 1 Fig. 17: Connection compartment, double chamber housing 1 2 3 4 Voltage supply, signal output For display and adjustment module or interface adapter For external display and adjustment unit Ground terminal for connection of the cable screen Information: The use of an external display and adjustment unit and a display and adjustment module in parallel in the connection compartment is not supported. 5.
5 Connecting to power supply Connection compartment 4...20mA (+)1 2 2(-) 1 Fig. 19: Connection compartment, Ex-d double chamber housing 1 2 Voltage supply, signal output Ground terminal for connection of the cable screen DIS-ADAPT with plug connector M12 x 1 for external display and adjustment unit 1 2 3 Fig. 20: View to the electronics compartment 1 2 3 DIS-ADAPT Internal plug connection Plug connector M12 x 1 4 3 1 2 Fig.
5 Connecting to power supply Wire assignment, connection cable Contact pin Colour connection cable in the sensor Terminal, electronics module Pin 4 Black 8 5.6 Wiring plan - version IP 66/IP 68, 1 bar 1 2 Fig. 22: Wire assignment fix-connected connection cable 1 2 Supplementary electronics - Second current output brown (+) and blue (-) to power supply or to the processing system Shielding 5.
5 Connecting to power supply SIM-Card Status Test USB (+)1 2(-) 1 Fig. 24: Radio module PLICSMOBILE integrated in the connection compartment 1 Voltage supply You can find detailed information on connection in the supplementary instructions "PLICSMOBILE GSM/GPRS radio module". 5.8 Switch-on phase After connecting the instrument to power supply or after a voltage recurrence, the instrument carries out a self-check for approx.
6 Functional safety (SIL) 6 Functional safety (SIL) 6.1 Objective In case of dangerous outages, processing facilities and machines can cause risks for persons, environment and property. The risk of such outages must be judged by the plant operator. Dependent thereon are measures for risk reduction through error prevention, error detection and fault control. The part of plant security depending on the correct functioning of safety-related components for risk reduction is called functional safety.
6 Functional safety (SIL) • Tool for operation and parameterization 4 … 20 mA current output 6.4 Safety concept of the parameterization The following tools are permitted for parameterization of the safety function: • • The integrated display and adjustment unit for on-site adjustment The DTM suitable for the signal conditioning instrument in conjunction with an adjustment software according to the FDT/DTM standard, e. g. PACTware Note: For operation of the VEGAFLEX 83 the DTM Collection version 1.
6 Functional safety (SIL) Change parameter All parameters changed by the operator are automatically stored temporarily so that they can be verified in the next step. Verify parameters/Lock adjustment After setup, the modified parameters must be verified (confirm the correctness of the parameters). To do this, you first have to enter the PIN. Here the adjustment is locked automatically. Then you carry out a comparison of two character strings. You must confirm that the character strings are identical.
6 Functional safety (SIL) Setup is then finished. Function test required Should a function test be necessary, the following message is displayed on the display and adjustment module. The adjustment software also signals that a function test is required. If a function test is required, the switching point or the range must be controlled with the original medium. For this purpose, you have to decide for your application which condition is potentially critical.
6 Functional safety (SIL) Measure the current output and compare the value with the calculated current value. You have to determine the permissible deviation of the values yourself. This deviation depends on the the accuracy requirements of your measurement loop. Determine the permissible tolerance for the deviation. Mode "Range monitoring" If both levels are important for the safety function, you have to proceed according to the mode "Range monitoring". 1.
7 Set up with the display and adjustment module 7 7.1 Set up with the display and adjustment module Insert display and adjustment module The display and adjustment module can be inserted into the sensor and removed any time. Four positions displaced by 90° can be selected. It is not necessary to interrupt the power supply. Proceed as follows: 1. Unscrew the housing cover 2. Place the display and adjustment module in the requested position onto the electronics and turn to the right until it snaps in 3.
7 Set up with the display and adjustment module 1 2 Fig. 26: Insertion of the display and adjustment module into the double chamber housing 1 2 In the electronics compartment In the connection compartment (with Ex-d-ia version not possible) Note: If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication, a higher cover with an inspection glass is required. 7.2 Adjustment system 1 44225-EN-130823 2 Fig.
7 Set up with the display and adjustment module • • • –– Confirm selected menu –– Edit parameter –– Save value [->] key: –– Presentation, change measured value –– Select list entry –– Select editing position [+] key: –– Change value of the parameter [ESC] key: –– Interrupt input –– Jump to next higher menu Adjustment system The sensor is adjusted via the four keys of the display and adjustment module. The LC display indicates the individual menu items.
7 Set up with the display and adjustment module Setup: Settings, e.g. measurement loop name, medium, vessel, adjustment, signal output, device unit, false signal suppression, linearization curve Display: Settings, e.g., for language, measured value display, lighting Diagnosis: Information, e.g.
7 Set up with the display and adjustment module correctly. If necessary, other measures must be taken to maintain the safety function. Setup - Measurement loop name Here you can assign a suitable measurement loop name. Push the "OK" key to start the processing. With the "+" key you change the sign and with the "->" key you jump to the next position. You can enter names with max. 19 characters.
7 Set up with the display and adjustment module Note: The selection of the application has a considerable influence on all other menu items. Keep in mind that as you continue with the parameter adjustment, individual menu items are only optionally available. You have the option of choosing the demonstration mode. In this mode, the sensor ignores the parameters of the application and reacts immediately to any change.
7 Set up with the display and adjustment module You can enter the dielectric constant of the upper medium directly or have the value determined by the instrument. To do this you have to enter the measured or known distance to the interface. Setup - Max. adjustment Level In this menu item you can enter the max. adjustment for the level. With interface measurement this is the maximum total level. Adjust the requested percentage value with [+] and store with [OK].
7 Set up with the display and adjustment module Setup - Max. adjustment - This menu item is only available if you have selected interface measInterface urement under the menu item "Application". You can accept the adjustment of the level measurement also for the interface measurement. If you select "Yes", the current setting will be displayed. If you have selected "No", you can enter the adjustment for the interface separately. Enter the requested percentage value.
7 Set up with the display and adjustment module The default setting is a damping of 0 s. Setup - Linearization A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e.g. a horizontal cylindrical or spherical tank, when the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume.
D -h +h 7 Set up with the display and adjustment module Fig. 28: Vessel height und socket correction value D Vessel height +h Positive socket correction value -h Negative socket correction value Since scaling is very extensive, scaling of the level value was divided into two menu items. Setup - Scaling Level 1 In menu item "Level 1" you define the scaling variable and the scaling unit for the level value on the display, e.g. volume in l.
7 Set up with the display and adjustment module Setup - Scaling Level 2 Since scaling is very extensive, scaling of the level value was divided into two menu items. In menu item "Level 2" you define the scaling format on the display and the scaling of the measured level value for 0 % and 100 %. Setup/Scaling Interface Since scaling is very extensive, scaling of the interface value was divided into two menu items.
7 Set up with the display and adjustment module Setup - Current output mode In the menu item "Current output mode" you determine the output characteristics and reaction of the current output in case of failure. The default setting is output characteristics 4 … 20 mA, failure mode < 3.6 mA. Setup - Current output Min./Max. In the menu item "Current output Min./Max.", you determine the reaction of the current output during operation. The default setting is min. current 3.8 mA and max. current 20.5 mA.
7 Set up with the display and adjustment module In the menu "HART variables 1" you can have the first HART value PV (primary value) and the second HART value SV (secondary value) of the sensor displayed. The values cannot be changed in the display and adjustment module. To change these values, you have to use the adjustment software PACTware. HART functions are not relevant for SIL applications.
7 Set up with the display and adjustment module Note: Check the distance to the product surface, because if an incorrect (too large) value is entered, the existing level will be saved as a false echo. The filling level would then no longer be detectable in this area.
7 Set up with the display and adjustment module In this case, you have to carry out a function test. Function test During a function test, you have to test the safety function of the instrument in the vessel with the original medium. For this purpose, you should know the filling height of the vessel as well as the min. and max. levels respectively for 4 and 20 mA. You then can calculate the respective output current.
7 Set up with the display and adjustment module To interrupt the function test, you have to push the button "ESC". If you carry out the function test by means of the "PACTware" software, you can store the previously performed tests and continue from there later on. Display In the main menu point "Display", the individual submenu points should be selected subsequently and provided with the correct parameters to ensure the optimum adjustment of the display options.
7 Set up with the display and adjustment module The lighting is switched off in the delivery status. Diagnostics - Device status In this menu item, the device status is displayed. Diagnostics - Peak values The respective min. and max. measured value is saved in the senDistance sor. The two values are displayed in the menu item "Peak values, distance".
7 Set up with the display and adjustment module Diagnostics - Peak values The respective min. and max. measured values are saved in the Additional sensor. The values are displayed in the menu item "Peak values Additional". This menu item displays the peak values of the electronics temperature as well as the dielectric constant. In another window you can carry out a reset of the two peak values separately.
7 Set up with the display and adjustment module Caution: During simulation, the simulated value is outputted as 4 … 20 mA current value and digital HART signal. Push the [ESC] key to deactivate the simulation. Information: The simulation is terminated automatically 60 minutes after the last key has been pushed. Diagnostics - Echo curve memory With the menu item "Setup" the echo curve it is possible to save at the time of setup.
7 Set up with the display and adjustment module You can find detailed information on the proof test in the Safety Manual (SIL). Additional settings - PIN Entering a 4-digit PIN protects the sensor data against unauthorized access and unintentional modification. In this menu item, the PIN is displayed or edited and changed. However, this menu item is only available if adjustment is enabled in the menu "Lock/Release setup/ adjustment". In delivery status, the PIN is "0000".
7 Set up with the display and adjustment module Setup Menu item Default value Lock adjustment Locked Measurement loop name Sensor Units Distance unit: mm Probe length Length of the probe Ex factory Medium type Liquid Application Level, vessel Medium, dielectric constant Water-based, > 10 Superimposed gas phase Yes Dielectric constant, upper medium (TS) 1.5 Tube inner diameter 200 mm Max. adjustment - Level 100 % Max. adjustment - Level Distance: 0.
7 Set up with the display and adjustment module Menu item Default value Scaling interface - 100 % correspond to 100 Scaling interface - 0 % correspond to 0 Current output, output variable Lin. percent - Level Current output - Output characteristics 0 … 100 % correspond to 4 … 20 mA Current output - Reaction in case of failure ≤ 3.6 mA Current output - Min. 3.8 mA Current output - Max. 20.
7 Set up with the display and adjustment module Menu item Default value Device memory - Measured value memory - Start with measured value Not active Device memory - Measured value memory - Stop with measured value Not active Device memory - Measured value memory - Stop recording when memory is full Not active Modified value Additional adjustments Menu item Default value PIN 0000 Date Actual date Time Actual time Time - Format 24 hours Probe type Device-specific HART mode Analogue curre
7 Set up with the display and adjustment module Additional adjustments Special parameters In this menu item you gain access to the protected area where you can enter special parameters. In exceptional cases, individual parameters can be modified in order to adapt the sensor to special requirements. Change the settings of the special parameters only after having contacted our service staff. Additional adjustments HART mode The sensor is fix adjusted to the HART mode "Analogue current output".
7 Set up with the display and adjustment module 7.4 Saving the parameter adjustment data We recommended noting the adjusted data, e.g. in this operating instructions manual, and archiving them afterwards. They are thus available for multiple use or service purposes. If the instrument is equipped with a display and adjustment module, the data in the sensor can be saved in the display and adjustment module.
8 Setup with PACTware 8 Via the interface adapter directly on the sensor Setup with PACTware 8.1 Connect the PC 2 1 3 Fig. 29: Connection of the PC directly to the sensor via the interface adapter 1 2 3 USB cable to the PC Interface adapter VEGACONNECT Sensor Via the interface adapter and HART 2 4 USB N OPE 3 TWIST LO CK 1 5 Fig.
8 Setup with PACTware such cases, the interface converter can be connected parallel to the 4 … 20 mA cable (dashed line in the previous illustration). Prerequisites 8.2 Parameter adjustment with PACTware For parameter adjustment of the sensor via a Windows PC, the configuration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required. The up-to-date PACTware version as well as all available DTMs are compiled in a DTM Collection.
8 Setup with PACTware In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves. The standard version is available as a download under www.vega.com/downloads and "Software". The full version is available on CD from the agency serving you. 8.
9 Set up with other systems 9 9.1 Set up with other systems DD adjustment programs Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™ and PDM. The files can be downloaded at www.vega.com/downloads under "Software". 9.2 Field Communicator 375, 475 Device descriptions for the instrument are available as EDD for parameter adjustment with the Field Communicator 375 or 475.
10 Diagnostics and service 10 Diagnostics and service 10.1 Maintenance If the device is used correctly, no maintenance is required in normal operation. When used in safety-instrumented systems (SIS), the safety function must be carried out on the instrument in regular time intervals by means of a proof test. Hence possible undetected, dangerous failure can be identified. The operator's responsibility to select the kind of test. The time intervals depend on the used PFDAVG.
10 Diagnostics and service • • • Switch-on and switch-off times Status messages (according to NE 107) Error messages (according to NE 107) The data are read out via a PC with PACTware/DTM or the control system with EDD. Echo curve memory The echo curves are stored with date and time and the corresponding echo data. The memory is divided into two sections: Echo curve of the setup: This is used as reference echo curve for the measurement conditions during setup.
10 Diagnostics and service Failure: Due to a malfunction in the instrument, a failure message is outputted. This status message is always active. It cannot be deactivated by the user. Function check: The instrument is in operation, the measured value is temporarily invalid (for example during simulation). This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
10 Diagnostics and service Code Cause Rectification F080 –– General software error –– Separate operating voltage briefly F105 –– The instrument is still in the start phase, the measured value could not yet be determined –– Wait for the end of the switch-on phase –– Duration depending on the version and parameter adjustment max. 5 min.
10 Diagnostics and service Code Cause Rectification C700 –– A simulation is active –– Finish simulation –– Wait for the automatic end after 60 mins. C701 –– Parameter verification was interrupted –– Finish parameter verification Text message Simulation active Parameter verification Out of specification The following table shows the error codes and text messages in the status message "Out of specification" and provides information on causes as well as corrective measures.
10 Diagnostics and service Code Cause Rectification M503 Reliability too low –– The echot/noise ratio is the small for a reliable measurement –– Check installation and process conditions –– Clean the antenna –– Change polarisation direction –– Use instrument with higher sensitivity M504 –– Hardware defect –– Check connections –– Exchanging the electronics –– Send instrument for repair M505 –– Level echo can no longer be detected –– Clean the antenna –– Use a more suitable antenna/sensor –– Remov
10 Diagnostics and service Error Cause –– Check connection according to chapter "Connection steps" and if necessary, correct according to chapter "Wiring plan" –– Voltage supply missing –– Check cables for breaks; repair if necessary –– Operating voltage too low or load resistance too high –– Check, adapt if necessary –– Electronics module in the sensor defective –– Exchange the instrument or send it in for repair Current signal greater than 22 mA or less than 3.
10 Diagnostics and service Measurement error with constant level 1. Measured value shows a too low or too high level Error pattern Level Fault description 0 Level 2. Measured value jumps towards 100 % time 0 time Cause Rectification –– Min./max. adjustment not correct –– Adapt min./max.
10 Diagnostics and service Measurement error during emptying 8. Measured value remains reproducible in one position during emptying Error pattern 0 Cause time –– Stored false signals in this position are larger than the level echo 0 Rectification –– False echo larger than the level –– Eliminate false signals in the echo close range –– Level echo too small –– Remove contamination on the probe. After having removed the source of the false signals, the false signal suppression must be deleted.
10 Diagnostics and service After an electronics exchange the settings of the electronics must be checked and verified. Only then, the instrument is again ready for operation. Exchanging the rod 10.6 Exchanging the rod The rod (meas. part) of the probe can be shortened, if necessary. To loosen the meas. rod you need a fork spanner with spanner width 10. Caution: Remember that the polished rod of the food version is very sensitive to damage and scratching.
10 Diagnostics and service Fig. 42: Exchanging the measuring rod Information: Please keep the stated torque so that the max. tensile strength remains. 7. Enter new probe length and if necessary the new probe type and then carry out a fresh adjustment (see "Setup procedure, Carrying out min. adjustment - Carrying out max. adjustment"). 10.
10 Diagnostics and service Make sure that you are using the correct software with SIL qualification. Instruments with SIL qualification can only be updated with a respective software. An accidental update with a wrong software version is impossible. Caution: Instruments with approvals can be bound to certain software versions. Therefore make sure that the approval remains effective with a software update. You can find detailed information on www.vega.com/downloads and "Approvals". 10.
11 Dismounting 11 Dismounting 11.1 Dismounting steps Warning: Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel or pipeline, high temperatures, corrosive or toxic products etc. Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order. 11.2 Disposal The instrument consists of materials which can be recycled by specialised recycling companies.
12 Supplement 12 Supplement 12.1 Technical data General data 316L corresponds to 1.4404 or 1.4435 Materials, wetted parts ƲƲ Process fitting 316L and PEEK ƲƲ Process seal On site ƲƲ Process seal on the instrument side (rod leadthrough) ƲƲ Rod: ø 8 mm (0.315 in) - polished Surface quality1) ƲƲ Polished (Basle Standard) ƲƲ Electropolished (Basle Standard) Materials, non-wetted parts FFKM (Kalrez 6221), EPDM (Freudenberg 70 EPDM 291) 316L (only 1.4435) - according to Basle Standard Ra < 0.
12 Supplement Torque for exchangeable rod probe (in the process fitting) 4.5 Nm ±0.5 (3.32 lbf ft) Torque for NPT cable glands and Conduit tubes ƲƲ Plastic housing ƲƲ Aluminium/Stainless steel housing Input variable Measured variable Min. dielectric constant of the medium max. 10 Nm (7.376 lbf ft) max. 50 Nm (36.88 lbf ft) Level of liquids εr ≥ 1.6 1 4 2 3 44225-EN-130823 5 Fig.
12 Supplement Output variable Output signal 4 … 20 mA/HART Fulfilled HART specification 7 Range of the output signal Signal resolution 3.8 … 20.5 mA/HART (default setting) 0.3 µA Failure signal current output (adjustable) ≥ 21 mA, ≤ 3.6 mA Max. output current 21.5 mA Load see load diagram under Power supply Starting current Damping (63 % of the input variable), adjustable ≤ 10 mA for 5 ms after switching on, ≤ 3.
12 Supplement ƲƲ Relative humidity 45 … 75 % ƲƲ Air pressure Installation reference conditions +860 … +1060 mbar/+86 … +106 kPa (+12.5 … +15.4 psig) ƲƲ Min. distance to installations > 500 mm (19.69 in) ƲƲ Medium Water/Oil (dielectric constant ~2.0)5) ƲƲ Vessel ƲƲ Installation Sensor parameter adjustment Typical deviation - Interface measurement Typical deviation - Total level interface measurement metallic, ø 1 m (3.
12 Supplement L 15 mm (0.591") 2 mm (0.079") 0 -2 mm (-0.079") -10 mm (-0.394") -15 mm (-0.591") 1 1 0,15 m (5.906") 0,35 m (13.78") 0,05 m (1.969") 0,1 m (3.937") Fig.
12 Supplement Gas phase Air Hydrogen Steam (saturated steam) Temperature Pressure 1 bar (14.5 psig) 10 bar (145 psig) 50 bar (725 psig) 20 °C/68 °F 0.00 % 0.22 % 1.2 % 200 °C/392 °F -0.01 % 0.13 % 0.74 % 400 °C/752 °F -0.02 % 0.08 % 0.52 % 20 °C/68 °F -0.01 % 0.10 % 0.61 % 200 °C/392 °F -0.02 % 0.05 % 0.37 % 400 °C/752 °F -0.02 % 0.03 % 0.25 % 100 °C/212 °F 0.26 % - - 180 °C/356 °F 0.17 % 2.1 % - 264 °C/507 °F 0.12 % 1.44 % 9.2 % 366 °C/691 °F 0.07 % 1.01 % 5.
12 Supplement 1 80°C (176°F) 2 0°C (32°F) -20°C (-68°F) 150°C (302°F) -40°C (-104°F) Fig.
12 Supplement Electromechanical data - version IP 66/IP 68 (1 bar) Options of the cable entry ƲƲ Cable gland ƲƲ Cable diameter (options) ƲƲ Cable entry ƲƲ Blind plug Connection cable 5 … 9 mm (0.20 … 0.35 in) 6 … 12 mm (0.24 … 0.47 in) 10 … 14 mm (0.39 … 0.55 in) ½ NPT M20 x 1.5; ½ NPT ƲƲ Wire cross-section 0.5 mm² (AWG 20) ƲƲ Tensile strength < 1200 N (270 lbf) ƲƲ Wire resistance ƲƲ Standard length ƲƲ Max. length ƲƲ Min. bending radius ƲƲ Diameter approx.
12 Supplement Voltage supply Operating voltage ƲƲ Non-Ex instrument, Ex-d instrument 9.6 … 35 V DC ƲƲ Ex-d-ia instrument 15 … 35 V DC ƲƲ Ex-ia instrument 9.6 … 30 V DC Operating voltage with illuminated display and adjustment module ƲƲ Non-Ex instrument, Ex-d instrument 16 … 35 V DC ƲƲ Ex-d-ia instrument 20 … 35 V DC ƲƲ Ex-ia instrument Interpolation protection 16 … 30 V DC Integrated Permissible residual ripple - Non-Ex, Ex-ia instrument ƲƲ for 9.6 V< UN < 14 V ƲƲ for 18 V< UN < 36 V ≤ 0.
12 Supplement Plastic housing ~ 69 mm (2.72") ø 79 mm (3.11") ~ 84 mm (3.31") ø 79 mm (3.11") M20x1,5/ ½ NPT 112 mm (4.41") 112 mm (4.41") M16x1,5 M20x1,5/ ½ NPT 1 2 Fig. 47: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher 1 2 Single chamber version Double chamber version Aluminium housing ~ 116 mm (4.57") ~ 87 mm (3.43") ø 86 mm (3.39") ø 86 mm (3.39") 120 mm (4.72") 116 mm (4.
12 Supplement Aluminium housing in protection rating IP 66/IP 68 (1 bar) ~ 105 mm (4.13") ~ 150 mm (5.91") ø 84 mm (3.31") ø 84 mm (3.31") 120 mm (4.72") 116 mm (4.57") M16x1,5 M20x1,5 M20x1,5 M20x1,5/ ½ NPT 1 2 Fig. 49: Housing versions in protection IP␣66/IP␣68 (1␣bar) - with integrated display and adjustment module the housing is 9␣mm/0.35␣in higher 1 2 Single chamber version Double chamber version Stainless steel housing ~ 87 mm (3.43") ~ 69 mm (2.72") ø 79 mm (3.11") ~ 59 mm (2.
12 Supplement Stainless steel housing in protection rating IP 66/IP 68, 1 bar ~ 93 mm (3.66") ~ 103 mm (4.06") ø 80 mm (3.15") ~ 105 mm (4.13") ø 79 mm (3.11") ø 86 mm (3.39") M20x1,5/ ½ NPT 120 mm (4.72") 112 mm (4.41") 117 mm (4.61") M16x1,5 M20x1,5 1 M20x1,5/ ½ NPT 2 3 Fig. 51: Housing versions in protection IP␣66/IP␣68 (1␣bar) - with integrated display and adjustment module the housing is 9␣mm/0.
12 Supplement 2 65,3 mm (2.57") ø 54 mm (2.13") 112 mm (4.84") 183 mm (7.20") VEGAFLEX 83, rod version ø 8 mm (0.315 in), polished - autoclaved version 3 1 L øw øw DIN DN25 DN32 DN40 / 1" 1 1/2" DIN DN50 / 2" DIN DN65 / 3" ø 50,5 ø 64 ø 91 Fig. 53: VEGAFLEX 83, rod version ø 8 mm (0.
12 Supplement 12.3 Industrial property rights VEGA product lines are global protected by industrial property rights. Further information see www.vega.com. Only in U.S.A.: Further information see patent label at the sensor housing. VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte. Nähere Informationen unter www.vega.com. Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle. Pour plus d'informations, on pourra se référer au site www.
INDEX INDEX A Accessories ––Display and adjustment module 14 ––External display and adjustment unit 15 ––Flanges 15 ––Interface adapter 14 ––Protective cap 15 Adjustment ––Max. adjustment 40, 41 ––Min. adjustment 40, 41 Adjustment system 36 Application 38, 39 Application area 11 B Backlight 49 Block adjustment 37 C Cable entry 21 Check signal 68 Connection cable 21 Connection procedure 22 Connection technology 22 Copy sensor settings 56 Current output 2 45 Current output Min./Max.
INDEX Probe length 38 Probe type 56 Proof test 52 R Read out info 57 Release adjustment 47 Repair 74 Replacement parts ––Display and adjustment module with heating 15 ––Electronics module 15 ––Spacer 15 Reset 53 S Scaling measured value 43, 44 Sensor characteristics 57 Sensor status 50 Service hotline 71 Shielding 21 SIL 29 Simulation 51 Special parameters 57 Status messages 64 Storage 14 T Type plate 10 U Units 38 V 44225-EN-130823 Verify parameter 33 Voltage supply 21 VEGAFLEX 83 • 4 … 20 mA/
All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice © VEGA Grieshaber KG, Schiltach/Germany 2013 VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.