Operating Instructions VEGASON 64 Foundation Fieldbus Ultrasonic
Contents Contents 1 About this document 1.1 1.2 1.3 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. .. .. .. .. .. .. .. 5 5 5 5 6 6 6 6 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. .. .. 8 9 10 10 General instructions . . . . . . . . . . . . . . . . . . . . . . . . . Mounting preparations . . . . . . . . . . . . . . . . . . . . . . . Mounting instructions . . . . . . . . . . . . . . . . . . . . . . . . 11 13 18 Configuration . . . . . . . . . . . . . . .
Contents 8 Maintenance and fault rectification 8.1 8.2 8.3 8.4 8.5 9 .. .. .. .. .. 42 42 43 44 44 Dismounting steps . . . . . . . . . . . . . . . . . . . . . . . . . . Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 46 Maintenance . . . . . . . . . . Remove interferences . . . . Exchanging the electronics Software update . . . . . . . . Instrument repair . . . . . . . ...... ...... module ...... ...... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 About this document 1 About this document 1.1 Function This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectification. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device. 1.2 Target group This operating instructions manual is directed to trained qualified personnel.
2 For your safety 2 For your safety 2.1 Authorised personnel All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator. During work on and with the device the required personal protective equipment must always be worn. 2.2 Appropriate use VEGASON 64 is a sensor for continuous level measurement. You can find detailed information on the application range in chapter "Product description".
2 For your safety 2.5 Safety approval markings and safety tips The safety approval markings and safety tips on the device must be observed. 2.6 CE conformity This device fulfills the legal requirements of the applicable EC guidelines. By attaching the CE mark, VEGA provides a confirmation of successful testing. You can find the CE conformity declaration in the download area of www.vega.com. 2.7 Fulfillment of NAMUR recommendations With respect to compatibility, the NAMUR recommendation NE 53 is fulfilled.
2 For your safety Chapter "Packaging, transport and storage" Chapter "Disposal" 28793-EN-081127 l l VEGASON 64 • Foundation Fieldbus 7
3 Product description 3 Product description 3.
3 Product description l Technical data: For example approvals, process temperature, process fitting/material, signal output, power supply, protection With the serial number, you can access the delivery data of the instrument via www.vega.com, "VEGA Tools" and "serial number search". In addition to the type label outside, you can also find the serial number on the inside of the instrument. 3.2 Principle of operation Application area VEGASON 64 is an ultrasonic sensor for continuous level measurement.
3 Product description 3.3 Operation VEGASON 64 can be adjusted with different adjustment media: l l l with indicating and adjustment module with the suitable VEGA DTM in conjunction with an adjustment software according to the FDT/DTM standard, e.g. PACTware and PC a configuration tool The entered parameters are generally saved in VEGASON 64, optionally also in the indicating and adjustment module or in PACTware. 3.
4 Mounting 4 Mounting 4.1 General instructions Suitability for process conditions Make sure that all parts of the instrument in contact with the measured product, especially the sensor element, process seal and process fitting, are suitable for the existing process conditions such as process pressure, process temperature as well as the chemical properties of the medium. You can find the specifications in chapter "Technical data" in the or on the type label.
4 Mounting Measuring range The reference plane for the measuring range (version A) is the lower edge of the flange. The reference plane for versions B, C and D is the lower edge of the transducer. Remember that with all versions there is a minimum distance below the reference plane - the so-called dead band - in which measurement is not possible; this distance from reference plane to medium must not be underrun. The exact value of the dead band is stated in chapter "Technical data".
4 Mounting 1 2 3 4 100% 0% Fig. 4: Version B, C, D - Measuring range (operating range) and max. measuring distance 1 full (dead band) 2 empty (max. measuring distance) 3 Measuring range 4 Reference plane Pressure/Vacuum Gauge pressure in the vessel does not influence VEGASON 64. Low pressure or vacuum does, however, damp the ultrasonic pulses. This influences the measuring result, particularly if the level is very low. With pressures under -0.
4 Mounting 3 Insert the transducer tube in the requested length from below into the swivelling holder 4 Fasten with screws (4); torque max. 10 Nm 5 Remove the plug from below out of the instrument housing and plug it into the socket of the transducer tube 6 Plug the electronics housing to the transducer tube. Do not squeeze any cables.
4 Mounting Mounting - Version C Version C is supplied in three parts: l l l Transducer Transducer cable electronics for wall mounting Assemble the version as follows: 1 Loosen the screws on the fixing ring (4) with a hexagon spanner (size 4), remove the transducer tube out of the swivelling holder 2 Mount the flange 3 Insert the transducer tube in the requested length from below into the swivelling holder 4 Fasten with screws (4); torque max.
4 Mounting 11 Secure screw (3) with lacquer or similar to ensure permanent tightness of the swivelling holder 12 Rotate the connection head in such a way that the cable entry points downwards (moisture can drain off). Fasten the cylinder with screws (5); torque max. 5 Nm 13 Insert the plug at the end of the transducer cable into the socket on the remote electronics 1 2 Fig.
4 Mounting 5 Plug the connection head to the transducer tube. Do not squeeze any cables. The cylinder is in the correct position, if the lower notch (2) of the transducer tube is no longer visible 4 1 3 2 Fig.
4 Mounting 4.3 Mounting instructions Mounting position When mounting VEGASON 64, keep a distance of at least 500 mm (19.69 in) to the vessel wall. If the sensor is installed in the center of dished or round vessel tops, multiple echoes can arise. These can, however, be suppressed by an appropriate adjustment (see chapter "Setup"). If you cannot keep this distance you should carry out a false echo storage before setup. This applies mainly if buildup on the vessel wall is expected.
4 Mounting In vessels with conical bottom it can be advantageous to mount the sensor in the center of the vessel, as measurement is then possible down to the lowest point of the vessel bottom. Fig. 11: Vessel with conical bottom Socket Preferably the transducer should be mounted without socket flush to the vessel top. If the reflective properties of the medium are good, you can mount VEGASON 64 on a socket piece. The socket end should be smooth and burr-free, if possible also rounded.
4 Mounting Vessel installations such as for example, ladders, limit switches, heating spirals, struts etc. can cause false echoes that interfere with the useful echo. Make sure when planning your measuring site that the ultrasonic signals have a "clear view" to the measured product. In case of existing vessel installations, a false echo storage should be carried out during setup.
4 Mounting Material heaps Large material heaps are best measured with several instruments, which can be mounted on e.g. traverse cranes. For this type of application it is advantageous to orient the sensor perpendicular to the bulk solid surface. Fig. 15: Transducers on traverse crane Inflowing medium Do not mount the instruments in or above the filling stream. Make sure that you detect the product surface, not the inflowing product. Fig.
4 Mounting Air turbulences If there are strong air currents in the vessel, e.g. due to strong winds in outdoor installations or air turbulence, e.g. by cyclone extraction you should mount VEGASON 64 in a standpipe or use a different measuring principle, e.g. radar or guided radar (TDR). Heat fluctuations Strong heat fluctuations, e.g. caused by the sun, can cause measurement errors. In this case, you should use a sun shield. Fig.
5 Connecting to power supply 5 Connecting to power supply 5.1 Preparing the connection Note safety instructions Always keep in mind the following safety instructions: l l Connect only in the complete absence of line voltage If overvoltage surges are expected, overvoltage arresters should be installed according to Foundation Fieldbus specification Tip: We recommend VEGA overvoltage arrester B63-32.
5 Connecting to power supply In systems without potential equalisation, connect the cable screen directly to ground potential at the power supply unit and at the sensor. In the connection box or T-distributor, the screen of the short stub to the sensor must not be connected to ground potential or to another cable screen. The cable screens to the power supply unit and to the next distributor must be connected to each other and also connected to ground potential via a ceramic capacitor (e.g. 1 nF, 1500 V).
5 Connecting to power supply 3 Remove approx. 10 cm (4 in) of the cable mantle (current output), strip approx. 1 cm (0.
5 Connecting to power supply 5.3 Wiring plan, double chamber housing Housing overview 1 2 3 5 4 Fig. 1 2 3 4 5 20: Double chamber housing Housing cover, connection compartment Blind stopper or plug M12 x 1 for VEGADIS 61 (optional) Housing cover, electronics compartment Filter element for air pressure compensation Cable gland Electronics compartment 1 Display I²C 1 2 5 6 7 8 2 3 Fig.
5 Connecting to power supply Connection compartment 1 BUS 3 L1 N 1 2 3 4 2 Fig. 1 2 3 22: Connection compartment, double chamber housing Spring-loaded terminals for signal output Ground terminal for connection of the ground conductor and screen Spring-loaded terminals for voltage supply Wiring plan BUS L1 N 1 1 2 3 /L /N PE 4 2 Fig. 23: Wiring plan, double chamber housing 1 Power supply 2 Signal output 5.
5 Connecting to power supply Then the current measured value will be displayed and the corresponding digital output signal will be outputted to the cable.1) 28 The values correspond to the actual measured level as well as to the settings already carried out, e.g. default setting.
6 Set up with the indicating and adjustment module PLICSCOM 6 Set up with the indicating and adjustment module PLICSCOM 6.1 Short description Function/Configuration The indicating and adjustment module PLICSCOM is used for measured value display, adjustment and diagnosis.
6 Set up with the indicating and adjustment module PLICSCOM Fig. 24: Mounting the indicating and adjustment module Note: If you intend to retrofit the instrument with an indicating and adjustment module for continuous measured value indication, a higher cover with an inspection glass is required.
6 Set up with the indicating and adjustment module PLICSCOM 6.3 Adjustment system 2 1 1.
6 Set up with the indicating and adjustment module PLICSCOM 6.4 Setup procedure Parameter adjustment As VEGASON 64 is a distance measuring instrument, the distance from the sensor to the product surface is measured. To have the real product level displayed, an allocation of the measured distance to the percentage height must be made. To carry out this adjustment, the distance is entered with full and empty vessel. If these values are not known, an adjustment with the distance values, e.g.
6 Set up with the indicating and adjustment module PLICSCOM 5 Carrying out max. adjustment Save the settings with [OK] and move to "Max. adjustment" with [>]. Proceed as follows: Max. adjustment 100.00 % = 1.000 m(d) 2.000 m(d) Medium selection 1 Prepare the % value for editing with [OK] and set the cursor to the requested position with [->]. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
6 Set up with the indicating and adjustment module PLICSCOM Enter the requested parameter via the appropriate keys, save your settings and jump to the next menu item with the [->] key. Damping To suppress fluctuations in the measured value display, e. g. caused by an agitated product surface, a damping can be set. This time can be between 0 and 999 seconds. Keep in mind that the reaction time of the entire measurement will then be longer and the sensor will react to measured value changes with a delay.
6 Set up with the indicating and adjustment module PLICSCOM Proceed as follows: 1 Move from the measured value display to the main menu by pushing [OK]. 2 Select the menu item "Service" with [->] and confirm with [OK]. Now the menu item "False signal suppression" is displayed. 3 Confirm "False signal suppression - Change now" with [OK] and select in the below menu "Create new". Enter the actual distance from the sensor to the product surface.
6 Set up with the indicating and adjustment module PLICSCOM Copy sensor data Copy sensor data? Reset Basic adjustment If the "Reset" is carried out, the sensor resets the values of the following menu items to the reset values (see chart):2) Function Reset value Max. adjustment 0 m(d) Min. adjustment Meas.
6 Set up with the indicating and adjustment module PLICSCOM 6.5 Menu schematic Basic adjustment ▶ Basic adjustment Display Diagnostics Service Info 1 Min. adjustment 000.0 % = 10.000 m(d) 1.245 m(d) 1.1 Damping 1.5 0s Max. adjustment 100.0 % = 0.000 m(d) 6.789 m(d) 1.2 Linearisation curve 1.6 1.3 Medium Vessel form Bulk solid 1.4 Silo linear Display ▶ Basic adjustment Display Diagnostics Service Info Displayed value 2 2.
6 Set up with the indicating and adjustment module PLICSCOM Service ▶ Basic adjustment Display Diagnostics Service Info 4 Gating out of false signals 4.1 Change now? Extended setting 4.2 Fast 4.3 Simulation 4.4 Reset Start simulation? Reset now? level change (> 1 m/min.) Unit of measurement 4.5 Language 4.6 Deutsch m(d) Copy sensor data 4.7 Copy sensor data? 4.8 PIN Enable? Info ▶ Basic adjustment Display Diagnostics Service Info 5 Device ID 5.1 < max.
6 Set up with the indicating and adjustment module PLICSCOM 6.6 Saving the parameter adjustment data It is recommended noting the adjusted data, e.g. in this operating instructions manual and archive them afterwards. They are hence available for multiple use or service purposes. If VEGASON 64 is equipped with an indicating and adjustment module, the most important data can be read out of the sensor into indicating and adjustment module.
7 Setup with PACTware and other adjustment programs 7 Setup with PACTware and other adjustment programs 7.1 Connect the PC via VEGACONNECT Internal connection via I²C interface 2 1 3 Fig. 1 2 3 External connection via I²C interface 26: Connection of the PC via VEGACONNECT directly to the sensor USB cable to the PC VEGACONNECT Sensor 1 2 TWIST OP EN 40 4 27: Connection via I²C connection cable I²C bus (com.
7 Setup with PACTware and other adjustment programs Necessary components: l l l l VEGASON 64 PC with PACTware and suitable VEGA DTM VEGACONNECT Power supply unit or processing system 7.2 Parameter adjustment with PACTware Further setup steps are described in the operating instructions manual "DTM Collection/PACTware" attached to each CD and which can also be downloaded from our homepage. A detailed description is available in the online help of PACTware and the VEGA DTMs.
8 Maintenance and fault rectification 8 Maintenance and fault rectification 8.1 Maintenance When used in the correct way, no special maintenance is required in normal operation. 8.2 Remove interferences Reaction when malfunctions occur The operator of the system is responsible for taken suitable measures to remove interferences. Causes of malfunction A maximum of reliability is ensured. Nevertheless, faults can occur during operation. These may be caused by the following, e.g.
8 Maintenance and fault rectification In Ex applications, the regulations for the wiring of intrinsically safe circuits must be observed. Fault messages via the indicating/adjustment module ? E013 l no measured value available à sensor in boot phase à Sensor does not find an echo, e.g. due to faulty installation or wrong parameter adjustment ? E017 l Adjustment span too small à Carry out a fresh adjustment and increase the distance between min. and max.
8 Maintenance and fault rectification Information: When loading on site, first of all the order data must be downloaded from the Internet (see operating instructions manual "Oscillator"). Assignment The electronics modules are adapted to the respective sensor and distinguish also in the signal output or power supply. 8.
8 Maintenance and fault rectification Please ask the agency serving you for the address of your return shipment. You can find the respective agency on our website www.vega.
9 Dismounting 9 Dismounting 9.1 Dismounting steps Danger: The individual wires of the transducer cable carry a voltage of approx. 70 V during operation (see chapter "Mounting preparations"). Therefore dismount VEGASON 64 only in complete absence of line voltage. Warning: Before separating the transducer cable (see chapter "Mounting preparations", example see illustration below), idle condition must be provided. Non-observance can cause damage to the electronics! 1 2 3 Fig.
9 Dismounting Correct disposal avoids negative effects to persons and environment and ensures recycling of useful raw materials. Materials: see chapter "Technical data" 28793-EN-081127 If you have no possibility to dispose of the old instrument professionally, please contact us concerning return and disposal.
10 Supplement 10 Supplement 10.1 Technical data General data Materials, wetted parts - Flange PP or Aluminium - Swivelling holder, flange - Transducer PA (1.
10 Supplement Input variable Measured value - Version A - distance between lower edge of the flange and product surface version B, C and D distance between lower edge of the transducer and product surface Measuring range - Liquids - up to 25 m (82.02 ft) up to 15 m (49.21 ft) Bulk solids 1 m (3.
10 Supplement 50 mm (1.196 in) 25 mm (0.984 in) 6 mm (0.236 in) -6 mm (-0.236 in) 3m 5 m (16.4 ft) (9.48 ft) 10 m (32.8 ft) 15 m (49.213 ft) 20 m (65.617 ft) 25 m (82.021 ft) -25 mm (-0.984 in) -50 mm (-1.196 in) Fig. 29: Deviation VEGASON 64 version A 50 mm (1.196 in) 25 mm (0.984 in) 6 mm (0.236 in) -6 mm (-0.236 in) 3m 5 m (16.4 ft) (9.48 ft) 10 m (32.8 ft) 15 m (49.213 ft) 20 m (65.617 ft) 25 m (82.021 ft) -25 mm (-0.984 in) -50 mm (-1.196 in) Fig.
10 Supplement Ambient conditions Ambient, storage and transport temperature -40 … +80 °C (-40 … +176 °F) Process conditions Process pressure - Version A with PP flange - 0 kPa -20 … 50 kPa/-0.2 … 0.5 bar (-2.9 … 7.3 psi) Other versions Process temperature (transducer temperature) -40 … +80 °C (-40 … +176 °F) Vibration resistance mechanical vibration with 1 g in the frequency range 5 … 200 Hz9) Electromechanical data Cable entry - Double chamber housing l 1 x cable gland M20 x 1.
10 Supplement Materials - Housing - Inspection window ABS Polyester foil Power supply Operating voltage - Non-Ex and Ex-d instrument 20 … 72 V DC, 20 … 253 V AC, 50/60 Hz Power consumption max. 4 VA; 2.1 W Electrical protective measures Protection IP 66/IP 67 Overvoltage category III Protection class I Approvals Depending on the version, instruments with approvals can have different technical data. For these instruments, the corresponding approval documents have to be taken into account.
10 Supplement 10.2 Foundation Fieldbus Block diagram, measured value processing The following illustration shows the Transducer Block (TB) and Function block (FB) in simplified form.
10 Supplement Secondary_value_1 cal_level_hi cal_level_lo Calibration Calibration Lowest Point Highest Point Sensor_value Sensor reference plane cal_level_hi [%] Calibration Highest Point cal_level_lo [%] Calibration Lowest Point Fig. 32: Adjustment VEGASON 64 Parameter list for Device revision 3.0 The following list contains the most important parameters and their meaning: l l l l l l 54 primary_value_unit Unit code of 'Primary_value' secondary_value_1 Value after min./max.
10 Supplement l l l l l l l l l l l l l l l l l l l - sensor_range Sensor_range.
10 Supplement - Min./max.-adjustment: Level at 'Cal_point_lo'. When writing 'Cal_level_lo' and 'Cal_type' = 1 (Online), the 'Cal_point_lo' is automatically set to the current sensor value. The unit is defined in 'Level_unit' l CAL_TYPE Min./max.-adjustment: Defines type of calibration: Dry: no influence of sensor value.
10 Supplement - Distance from the sensor to the product surface. Unit derives from 'Sensor_range.unit' l empty_vessel_curve_corr_op_code Update, create new or delete the empty vessel curve l l l l l l sound_velocity_unit Unit code of 'Sound_velocity' Temperature Process temperature. Selected as input to AIFB by setting 'Channel' = 4. Unit derives from 'Temperature.unit' temperature_unit Unit code of 'Temperature', 'Max./Min.
10 Supplement 10.3 Dimensions Housing 27/ ~ ") ø 84mm 5/16") 120mm (4 23/32") * 129mm (5 5/64") M12x M20x1,5/ ½ NPT Fig.
10 Supplement VEGASON 64 130 mm (5.12") 1 max. 15° (0.59") b 440 mm (17.32") 240 mm (9.45") 2 110 mm (4.33") 3 4 D k ø 189 mm / ø 196 mm (7.44") (7.72") k D mm DN 50 DN 80 inch DN 50 / 2" DN 80 / 3" D 165 200 b 11,5 11,5 k 125 160 d 4x ø 18 4x ø 18 >ø 200 mm (7.87") 250 mm (9.84") DIN D b k d 6.50" 0.45" 4.92" 4x ø 0.71" 7.87" 0.45" 6.3" 4x ø 0.71" ASME mm 2" 3" inch 2" 3" D 6" 7.5" b 11,5 11,5 k 120,7 152,4 d 4x ø 19,1 4x ø 19,1 3 4 b k d 0.45" 4.75" 4x ø 0.75" 0.
10 Supplement VEGASON 64 150 mm (5.91") 2 k 143 mm (5.63") D 18 mm (0.71") 312 mm (12.28") 220 mm (8.66") 1 149 mm (5.87") G1A >ø 200 mm (7.87") 250 mm (9.84") 3 4 DIN mm DN 50 DN 80 D 165 200 b 11,5 11,5 D 152,4 190,5 b 11,5 11,5 k 125 160 d 4x ø 18 4x ø 18 inch DN 50 / 2" DN 80 / 3" D b k d 6.50" 0.45" 4.92" 4x ø 0.71" 7.87" 0.45" 6.3" 4x ø 0.71" ASME mm 2" 3" 60 inch 2" 3" D 6" 7.5" b k d 0.45" 4.75" 4x ø 0.75" 0.45" 6" 4x ø 0.
10 Supplement 10.4 Industrial property rights VEGA product lines are global protected by industrial property rights. Further information see http://www.vega.com. Only in U.S.A.: Further information see patent label at the sensor housing. VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte. Nähere Informationen unter http://www.vega.com. Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle.
10 Supplement 28793-EN-081127 62 VEGASON 64 • Foundation Fieldbus
28793-EN-081127 10 Supplement VEGASON 64 • Foundation Fieldbus 63
Printing date: VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info@de.vega.com www.vega.com ISO 9001 All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing systems correspond to the information available at the time of printing.