Setting the Low Glucose Suspend
If the Glucose, Predictive, and Fall Rate of Change alerts are set up, your pump will generate an alert to
warn you about low or dropping sensor glucose values. However, if you do not treat or are unaware of your
low blood glucose, you can set the Low Glucose Suspend feature to stop insulin delivery. When your current
sensor glucose reaches or drops below your Low Glucose Suspend limit, your pump stops insulin delivery, and
triggers the LOW SUSPEND alarm. See LOW SUSPEND, on page 190 for more information. The default setting
for the Low Suspend feature is off. Check with your healthcare professional to determine the Low Suspend
setting that is best for you.
To set up the Low Glucose Suspend:
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Low Suspend, then press ACT. The LOW SUSPEND screen appears.
3 Select On, then press ACT.
4 The SET LOW SUSPEND LIMIT screen flashes the default Low Suspend Limit of 3.3 mmol/L (60 mg/dL).
Select your limit. You should set the Low Suspend Limit to a value that is less than your Low Glucose
Limit. For example, if your Low Glucose Limit is set at 4.2 mmol/L (76 mg/dL), you should set your Low
Suspend Limit to a value that is less than 4.2 mmol/L (76 mg/dL). The Low Suspend Limit should be
between 2.2 and 6.1 mmol/L (40 to 110 mg/dL).
5 Press ACT. The EDIT SETTINGS screen shows the Low Suspend Limit value set. You are now ready to set
up the Cal Repeat.
Setting the Cal Repeat
After you receive and clear a METER BG NOW alert, the device will repeat the alert until you enter a new
meter blood glucose measurement. The Cal Repeat feature allows you to set how frequently you want the
alert to repeat after you clear it.
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Cal Repeat, then press ACT. The SET CAL REPEAT screen flashes the default Cal Repeat time of
30 minutes (0:30).
3 Set your Cal Repeat time between 5 minutes (0:05) and 1 hour (1:00).
4 Press ACT. The EDIT SETTINGS screen shows the Cal Repeat time you selected. You are now ready to set
your Cal Reminder.