User guide

To keep your insulin delivery suspended after clearing the alarm:
1 Press ACT with the Suspend option highlighted in the LOW SUSPEND/CHECK BG screen.
2 The SUSPEND screen appears, and the pump soon returns to the HOME screen.
3 Check your BG and treat as appropriate. After your blood glucose has returned to an acceptable
measurement, see the following section to restart delivering your pump's programmed basal.
NOTE: Two hours after you choose to keep your insulin delivery suspended, the pump automatically
resumes delivering insulin at the basal rate (standard, pattern or temporary) that is programmed to be
active for that time. The pump returns to the HOME screen as insulin delivery resumes. When you press
ACT from the HOME screen, the following message appears: Basal delivery is resumed. Check BG.
Press any key to cont. If your sensor glucose is still at or below the Low Suspend Limit, your LOW
SUSPEND alarm will repeat as frequently as you have set in the Low Repeat feature. See Setting the
Low Repeat, on page 124 for more information.
To resume insulin delivery after you suspend your pump:
1 Press ACT from the HOME screen when you are ready to restart your basal delivery.
2 The LOW SUSPEND/CHECK BG screen appears. Select Restart Basal, then press ACT.
3 The pump resumes delivering insulin at the basal rate (standard, pattern or temporary) that is
programmed to be active for that time. The following message appears as insulin delivery resumes: Basal
delivery is resumed. Check BG. Press any key to cont.
4 Press any button to continue using your pump. If your sensor glucose is still at or below the Low Suspend
Limit, your LOW SUSPEND alarm will repeat as frequently as you have set in the Low Repeat feature. See
Setting the Low Repeat, on page 124 for more information.
Troubleshooting, alerts and alarms 191