User guide
Testing the transmitter
The tester acts like a sensor. If you get sensor-related alerts, use the tester to make sure that the transmitter
is working.
Connecting the tester
1 Hold the transmitter and the tester as shown. Line up the flat side of the tester with the flat side of the
transmitter.
2 Push the tester into the transmitter until the flexible side arms of the tester click into the notches on
both sides of the transmitter.
3 Within 20 seconds, the green light on the transmitter will flash for about 10 seconds when properly
connected. The Sensor feature must be turned On on your pump. Go to the RECONNECT OLD SENSOR
screen to start your sensor.
Main > Sensor > Link to Sensor > Reconnect Old Sensor
You should see the sensor icon
on the pump screen, which means that the transmitter and the sensor
are communicating.
4 Within the next seven to eight minutes, go to the SENSOR STATUS screen to look for the Sensor ISIG value:
a. The sensor ISIG value on this screen should be between 24.00–29.00 nA. This range of ISIG values
means that the transmitter electronics are working properly. Since the transmitter is sending signals
correctly, the sensor must have caused the alert. Remove and discard the sensor. Insert a new sensor
in a new site.
b. If you see a Sensor ISIG value that is less than 24.00 nA or more than 29.00 nA, contact your local
representative. It may be time to replace your transmitter.
Troubleshooting, alerts and alarms192