Operating instructions

Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 25
Gas Supply Requirements
Important: To prevent damage to the detector, do not exceed the maximum gas
pressure of 550 kPa (80 psig) for the gas supply.
Important: Ne pas dépasser une pression de gaz maximaux de 550 kPa (80 psig)
pour l'approvisionnement en gaz.
If you work with combustible solvents, use nitrogen gas. Higher purity nitrogen is not
required but recommended and should be dry (not containing water vapor). Medical-
grade nitrogen from an in-house liquid nitrogen source is sufficient. However, filter the
supplied gases through a submicron particle filter before they enter the detector. Not
using a well-filtered operating gas supply may increase the baseline noise.
Observe the following notes:
The inlet gas must be clean and free from micro particles (size 0,1 µm).
The inlet gas supply pressure must be between 480 550 kPa (70 80 psig).
The gas consumption of the detector is approximately 4 L/min.
Pressure variations in which the gas supply drops below 480 kPa (70 psig) may result in
loss of performance. Monitor the gas supply pressure during the installation.