Operating instructions

Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 40 Operating Instructions
4.4.5 Connecting the Stream-Switching Valve (Corona Veo RS only)
To connect the stream-switching valve, connect ports 2, 3 and 4 of the valve. The ports of
the stream-switching valve are assigned as follows:
No. Ports
2 Divert/Waste (away from the
detector)
3 Inlet (capillary from column)
4 Nebulizer (to nebulizer inlet)
Fig. 6: Ports on the stream-switching valve
1. Make sure that you have connected the detector to the flow path as described in
section 4.4.4 before you proceed.
2. Locate the Viper capillary, the gray capillary and the RheFlex fittings from the
accessories kit for the detector.
3. Connect the Viper capillary from the Nebulizer port (Port 4) on the stream-switching
valve to the nebulizer inlet using the RheFlex fittings.
4. Connect the gray capillary to the Divert/Waste port (Port 2) using a RheFlex fitting
and direct it to the waste.