5.3.3.1 Diagnostics Screen
The Diagnostics screen displays information with regards to factory-set and internally
calibrated parameters. Real-time values are displayed.
Fig. 9: Diagnostics screen (here: Corona Veo RS)
The following functions and parameters are available on the screen:
Parameter Description Typical Values
Flow Ratio Indicates the ratio of nebulizer gas pressure to the
charging gas pressure calibrated at factory.
Normal
Ion Trap Removes highly mobile gas ions. 18 V 20.9 V
Gas Pressure Indicates inlet pressure from gas source. 276 448 kPa
(40-65.0 psig ± 0.1
psig)
Evaporation Temp Displays if the evaporation tube is operating and the
actual evaporation temperature.
35 C
Charger Voltage Indicates the applied voltage to the corona wire to
charge the (secondary) ion gas stream.
1.8 kV 3.0 kV
Charger Current Indicates the current required to charge the (secondary)
ion gas stream.
0.99 1.01 µA
To return to the Run Mode Menu, press the Settings soft key.
Flow Ratio = 0.42
Ion Trap = 20.5 V
Gas Pressure = 62.9 psi
Evaporation Temp = On, 35 C
Charger Voltage = 2.25 kV
Charger Current = 0.99 µA
Exit
Run Mode
Graph
Auto Zero
Evap T
DIAGNOSTICS