Operating instructions
Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 54 Operating Instructions
5.3.3.3 Evaporation Temperature Screen
The Evaporation Temperature screen provides the means to set and monitor the
temperature of the evaporation tube.
Fig. 11: Evaporation Temperature screen (here: Corona Veo RS)
The following functions and parameters are available on the screen:
Parameter Description
Temperature Displays the actual evaporation temperature.
Power Turns on and off temperature control of the evaporation tube.
Setpoint Sets the target temperature for the evaporation tube; default temperature is 35 °C.
Corona Veo: Two temperature settings are available: 35 °C and 50 °C.
Corona Veo RS: The temperature can be set from ambient +5°C to +100°C in 0.1°C
increments.
Tips: The actual temperature reached depends on several factors, including mobile
phase composition, liquid flow rate, mobile phase temperature, gas
temperature, and ambient temperature.
To ensure that the evaporation tube comes to equilibrium for the set
temperature, wait approximately 30 minutes after the temperature is changed
before you start sample analysis.
EVAPORATION TEMPERATURE
Temperature: 35 C
Setpoint: (35) C
Power: ( On )
Exit
Run Mode
Graph
Auto Zero
Diagnostics