User's Manual Part 1

FloPoint
®
Elite Uroflow System Getting Started
User’s Manual page 25
Part and Part Number Name and Description
0600-0232
Label Writer Power Cord
Connects the Label Writer power adaptor to a wall outlet.
0275-0002
Power Supply:
Connects the power cord to the Label Writer.
0125-0446
Roll of Labels:
Labels in roll format properly sized for the ScanPoint
®
Label
Writer.
Step 2: Attach the Mounting Bracket to the wall and attach the FloCharger to the
Mounting Bracket.
For an installation of the FloCharger in the patient vicinity to comply with IEC 60601-1
and the National Electric Code in the United States and the Canadian Electric Code
in Canada, no part of the FloCharger cord must be lower than 3 feet above the floor.
Use the cord wrap underneath the FloCharger to protect the unused portion of the
cord. Any other installation method will jeopardize compliance of the FloPoint
®
Elite
system with IEC 60601-1 and the NEC or CEC and will therefore void any warranties
(implied or otherwise) offered by Verathon Medical
®
for the instrument.
WARNING: Install the FloCharger Mounting Bracket at a height of at least 3 feet (10
meters) above the floor.
WARNING: Use the cord wrap underneath the FloCharger to protect the unused
portion of the cord.
WARNING: Installation of a FloCharger that does not use the Mounting Bracket or
alternate protection of the cord will not be compliant with the Medical-Electrical Safety
Standard (IEC 60601-1). Alternate protection requires that the cord be 3 feet (10
meters) above the floor and not easily accessible by a patient.