User's Manual Part 3

FloPoint
®
Elite Uroflow System Calibration
User’s Manual page 81
Calibration
To make sure the FloSensor
®
is providing accurate measurements, it must be calibrated
at prescribed intervals.
The entire calibration process takes approximately 15 minutes.
You can calibrate the FloSensor as often as you like; you do not have to wait for the next
scheduled calibration date. However, at a minimum, you must calibrate the FloSensor
every 12 months.
NOTE: If your facility has signed up for the ScanPoint
®
Online Service Plan, you will
automatically receive the Calibration Fluid 4-pack and a Calibration Quick Reference
Card approximately 30 days before your FloSensor requires calibration (Figure 26).
Figure 26. FloPoint
®
Elite Calibration Fluid and Calibration Quick Reference Card.
0800-0331
Calibration Fluid Pouch
Pouch containing unique fluid for use in calibrating the
FloPoint
®
Elite.
0900-1444
FloPoint
®
Elite Calibration Quick Reference Card:
Provides instructions for the FloPoint
®
Elite calibration
procedure.
Determining the Number of Days Until Required Calibration
1. Clear the Remote of saved exam data by placing it in the ScanPoint
®
Docking
Station and saving the exam data to ScanPoint
®
.
2. Remove the Remote from the ScanPoint
®
Docking Station. Press and hold down
the Start button for at least five seconds. The LCD screen displays the “setting
Calibration