Specifications
22-0375 rev1.1  Monitor ISM/xL™ Hardware Guide  25 
Graphic Map Annunciator Module 
Features: 
•  Recommended mounting height: 1.5 meters (4 feet, 9 inches). 
•  Can be assigned with maximum 16 outputs to turn on any 16 LEDs over the display of 70 possible LED locations. 
• This module can be assigned with 4 hardwire inputs. 
•  Outputs 1 and 2 can be the first 2 LEDs on the screen or used as hardwire outputs or both as these 2 outputs will 
turn on both sources (LEDs or hardwire connections) at the same time. 
•  When outputs 1 and 2 are used as hardwire connections they behave the same as on the input/output modules. 
Output 1 turns positive 12V and output 2 turns negative. 
•  The 0V negative for hardwire positive output #1 and 12V positive for hardwire negative output #2 can also be used 
as auxiliary 12V power connections (fused at .5 AMP). 
•  Applying a magnet to a reed switch on the PCB will turn on all LEDs in the display for testing. 
•  The Map Module drawing software is a Windows based draw program. It is installed as a template to use over and 
over. It allows a user to do a diagram of a system’s layout for the LEDs to turn on and highlight certain point 
programmable output conditions in the diagram. E.g. point not normal, point in alarm. It could also be used to make 
alpha/numeric labels that can have the LEDs turn on next to them for various area, point indications. E.g. “Area 1 
ON”, “Area 2 STAY”. When the diagram is complete, it can be printed out and cut out to fit inside the Map Module’s 
display. 
•  The Map Module drawing software is available on the Director Software CD and listed as “MAP MODULE 
TEMPLATE.dot.” 
D5.....D1
C10.....C1
B10.....B1
A10 A1
G10.....G1 F10.....F1 E10.....E1
D10...D6
1-
BROWN
2-
RED
4-
YELLOW
5-
GREEN
6-BLUE
7-
PURPLE
16-BLUE
WHITE
9-
8-GREY
10-
BLACK
11-BROWN
13-ORANGE
14-YELLOW
15-GREEN
12-
RED
3-ORANGE
RIBBON CABLE
Each ribbon 
cable coloured 
wire represents 
outputs 1 to 16. 
They are 
connected to the 
desired pin on the 
pin strips to turn 
on the required 
LED # on the 
face of the 
module to display 
a condition. E.g. 
follow an input 
that is not normal.
Data A
Data B
F1
1/2 AMP
P/N 342-3356
1432
312
COM
I/P3
I/P4
I/P1
(+) 12V
0V
Module Bus
red
green
yellow
black
TB1
Bus 
communications 
indicator LED 
312
LED PIN CONNECTION STRIP
LED PIN CONNECTION STRIP
Tamper 
Spring
(+)12V 
(Aux Pow)
1432
0V (Aux Pow)
COM
I/P2
O/P1
O/P2
Switched (+) 
Switched (-) 
TB9TB13
5 digit address to program 
in Module Group
LED Test Reed Switch
hold magnet over it and 
all LEDs on face will 
turn on.
TB2
Graphic Map Module
Current Rating = 150mA max when 
all LEDs on.
XXXXX
Refer to Installation Instructions 
P/N 22-0364 for further 
information.