Instruction manual
Albatros II
2012-09
Page 15
2.6 Emergency
In cases for emergency the emergency stop and the function for lowering in an emergency
must be used.
2.6.1 Emergency stop
In emergencies, and during assembly and disassembly,
the emergency stop is activated by pressing the red
button on the control box.
Push on the button and turn in the direction of the arrows.
If the button is fully pressed to the back turn back in the
opposite direction of the arrows.
The emergency stop is deactivated by
turning the red button in the direction
of the arrows.
2.6.2 Manual emergency lowering
L WARNING: Risk of injury – Adjust the manual emergency lowering for the weight
of the patient.
Emergency lowering if the power fails or the batteries are
empty can be done using the red pull button (1) at the
lower end of the motor. This emergency lowering is
factory-set for a patient weight of 50 kg.
Standard calibration: 2500N, 15 mm/s to lower the
patiënt.
Please note that the manual emergency lowering is only possible when the patient is sitting
in the patient lift.
Pull force: 10 mm for force from 80 à 100N.
1