Instructions / Assembly
Table Of Contents
Packing List
Serial
no.
category
Name
Unit
Qty
confirmation
*
1
documents
Operation Manual
piece
1
2
documents
Product inspection certificate
piece
1
3
documents
Warranty Card
piece
1
4
Host
Drying oven
set
1
5
Parts
Shelf
piece
2
CE
timing unit
0-1
0 takes a timing unit of minute, and 1 takes a timing unit of
hour.
0
Ct
Refrigeration
control delay
0
~
3600
秒
The time required for two adjacent compressor starts, Ct=0
means to cancel the compressor function.
0
CH
upper bias
setting of
refrigeration
control
0~full
scale
0.0~full
scale
When the temperature exceeds the CH value and conforms to
the refrigeration control delay of the compressor, the
refrigerating lamp lights on and refrigeration contact connects,
the compressor starts, namely SP+CH
〉
setting value
0.5
CL
Lower bias
setting of
refrigeration
control
0~full
scale
0.0
~
full
scale
When the compressor starts, the refrigeration contact
disconnects while the measured temperature
≤
setting
temperature +CH-CL value, the compressor stops.
0.3
drt
Open door
test
0-full
range;
scale span
When the temperature drops and the deviation exceeds the
deviation, the meter judges that the door has been opened
1.0