Instructions / Assembly

5
· Do not use the HOLD function to measure unknown voltage. After HOLD is
turned on, the display screen will not change when different voltages are
measured.
· Measure a known voltage to determine whether the instrument operates
normally.
· When measuring, the correct function and range gear must be used.
· Do not use the instrument around explosive gas and steam or in humid
environment.
· Do not use damaged probes. Check whether the insulation of the probe is
damaged, whether there is exposed metal or signs of wear. Check the continuity
of the probes.
· When measuring, please connect the neutral wire or ground wire first, and then
the live wire; when disconnecting, please cut off the live wire first, and then
disconnect the neutral wire and ground wire.
· When measuring, please hold your fingers behind the probe finger protector.
· Please disconnect the probe from the measured object before opening the back
cover of the instrument.
· Do not use the instrument in an environment that exceeds the measurement
category (CAT) rating of the lowest rated single element in the instrument, probe
or accessories.