User's Manual

Table Of Contents
Chapter 6 Maintenance and Troubleshooting
Calibrating the touchscreen
T-BERD/MTS 5800 Getting Started Manual
93
Before replacing the battery, verify that it is a JDSU-supplied
replacement battery pack.
To replace the battery pack
1 Turn the power off.
2 Turn the instrument over, resting on the face.
3 Using a phillips screw driver, remove the four screws from
the battery compartment door.
4 Remove the battery compartment door.
5 Pull the battery away from the connector on the left side.
6 Push the new battery into the connector on the left side.
7 Replace the battery compartment door.
8 Tighten the four screws until hand tight.
Replacing the battery pack is complete.
Calibrating the touchscreen
After some time, the touchscreen may need to be calibrated.
For example, maybe you touch the screen but the arrow
appears elsewhere, or it doesn’t recognize the very top or
bottom of the screen.
CAUTION: INSTRUMENT DAMAGE
Failure to turn the power off before removing the batteries
from the T-BERD MTS 5800 could damage internal compo-
nents and/or corrupt the software. Always power down the unit
before removing the battery pack.