CE300993 Tuning-Fork Scale HG S&J? ± O S ( CE TYPE ) Operation Manual -looo Tokyo
CONTENTS PAGE INTRODUCTION 1 Four Basic Scale Modes 1 Four Basic Functions 1 Thirteen Weight Unit Variation 2 Other Specifications 2 Model s 2 EXTERNAL VIEW & NAMES OF PARTS 3 Key Function 3 Characters 3 INSTALLATION 4 Unpacking, Unlocking, Pan Loading 4 Leveling, Starting 5 Cautions on Use 5 SCALE MODE SETTING 6 How to Select a SCale Mode 6 1. Precision Weighing Scale Mode 7 Taring, Bar Graph Display 7 2. Counting Scale Mode 8 3.
Part III Conditions & Results, Integration Time & Stabilization Speed Part IV Contents of INterface 23 23 LIST OF VARIOUS MESSAGES Displays, Beeps TROUBLESHOOTINGS CALIBRATION 24 24 25 26 BATTERY OPTION OUTPUT FORMAT 27 28 (2)
Congratulations to be one of the owners of the multi-purpose tuning-fork scale VIBRA - HG series ! With 6 different weight unit displays, the VIBRA - HG scale offers 16 varieties of function complex, which will satisfy you in any weighing field. To get the optimum result, select the most suitable function to your weighing purposes and conditions out of the full menu programmed for owners of VIBRA. Let's start to look into the world of VIBRA ! FOUR (4) BASIC SCALE MODES page 1.
SIX (6)WEIGHT UNIT VARIATION 1. mg 2. g 3. kg 4. ct 5. oz 6. lb -Two units out of these 6 weight units are selectable.- See page 21. OTHER SPECIFICATIONS 1. Measuring System 2. Taring Range 3. Zero Tracking 4. Span Adjustment 5. Overload Message 6. Temperature Coefficient 7. Temperature/ Humidity 8. Power Source 9. Display 10. Output 11. Functions 12. Options 13. Under Weighing Tuning-fork frequensy sensing system Full weighing range 4% of max.
EXTERNAL VIEW & NAMES OF PARTS Weighing Pan T key(Tare) F key(Function) S key (Setting Change) KEY FUNCTION AC Adaptor Setting Change & Finish Function/Mode change Tare & Parameter Change CHARACTERS g pcs % # M Weight unit in weighing, counting, percentage mode (mg, kg, ct repiacable) Unit indication in counting mode Percentage indication in percentage mode Unit indication in conversion mode Indication in accumulation function Indication that a particular setting has been stored.
INSTALLATION 1. UNPACKING Unpack- the container carefully. Examine the packaging and the device for damage, and report to the shipper if any. Try to keep the scale upright. Check for the exclosures as follows: (T) Scale ( D Operation Manual ( D Weighing Pan with Pan Base (?) AC Adaptor 2. UNLOCKING Locate a black rubber cap at the top of the scale. Remove the cap to find a lock lever in it. Turn the lever to the centre of the scale to unlock.
4. LEVELING Watch if the scale is level. Locate the level in left front of the scale, and knurled legs beneath the scale. Drive these legs to centre the bubble in the red circle of the level properly. LEVEL NOT LEVEL Level Red Circle Bubble Level Adjusters Level 5. GETTING STARTED (1) Connect the AC adaptor with the scale, then plug the cord in line outlet. (2) Turn on the1 power switch on the right side near bottom. The display will light up all segments and characters for about 4 sec. as a self test.
SCALE MODE SETTING FOUR SCALE MODES Select suitable scale mode to your purpose out of the following four modes. 1. Precision Weighing Mode : 1/250,000 to 1/150,000 divisions...see page 7. 2. Counting + Weighing Mode: High precision, easy counting see page 8. 3. Percentage+Weighing Mode: Percentage of weight/reference see page 10. 4. Conversion+Weighing Mode: Conversion by weight/coefficient...see page 12. * At the time of delivery, the scale mode is set at 1. Precision Weighing Mode.
OPERATIONS Regarding to measurement in other weight units, see page page 18 in 1. PRECISION WEIGHING SCALE MODE the column. You can select 2 units out of 6 each of Mode A and 8. WEIGHING WITHOUT TARE Connect the AC adaptor with the scale at the rear, plug the cord in line outlet. Turn the power switch of the scale. (1) Press the jT| key and verify that the display is cleared to "0". (2) Load objects on the scale and read the display.
2. COUNTING SCALE MODE The HG scale employs unique automatic memory renewal system, namely Simple Self Counting System. By this SCS, highly accurate reference unit weight will be obtained automatically, simply by adding optional number of samples (within three times of that the display requires ). COUNTING OPERATION BY SIMPLE SCS Tare/Zero setting 0 PCs (1) Place a container, if any, on the scale and press the jjj key. Verify that the display is "0".
(6) Add. some more samples, but within three times of the quantity displayed, to improve accuracy of the unit weight. Without any key operation, thus, the unit weight will be improved by increasing samples by and by repeatedly. (7) At any sample quantity you want to finish, pressing the j_Fj key will finish the improvement operation. A beep comes out and the display will turn to ordinary counting mode. Automatic unit weight improvement rvV .rVx )O BS pcs Improvement (sampling) finish g c O NOTES 1.
3. PERCENT SCALE MODE 100% SETTING WITH REFERENCE SAMPLE Percentage measurements by a reference sample as 100% memorizing in the scale. (1) Press the [ D key to indicate "0". Tare/Zero setting 0 Also press the [T] key when use a tare container. L (2) Press the |_FJ key until P SEh appears. M sign and some value will start blinking. The value shows 100% reference value currently set. \ (S o Reference setting SEt P. 0.
100% SETTING BY KEY OPERATION Percentage measurements per 100% reference value set by key operation. (1) Press the L L J key until P S E t appears. Reference setting p. set M sign and some value will start blinking. The value shows 100% reference value currently m set. n n n v u.u u A (2) Pressing the [73 key will display "0" and bar Manual setting start n v graphs under it blinking, to start manual setting m m of the reference value for 100%.
4. CONVERSION SCALE MODE In this scale mode, the scale displays a weight value of the load, multiplied by a coefficient memorized in the scale. With this function, the HG scale offers measurements in countless weighing units. OPERATION Coefficient setting (1) Press the J T ] key until C S E t appears. C. SEt M sign and some value will start blinking for setting new coefficient. m The value blinking. is the current coefficient in the memory. (2) Press the \Jj key.
VARIOUS FUNCTIONS FOUR BASIC FUNCTIONS The HG scale has Accumulation and Comparator function in addition to the high precision weighing function. In combination of these functions, it offers four basic function complex as follows: page A. Precision Weighing Function : Simple display of weight data 7 B. Accumulation Function : Display of sum of measured data... 15 C. Comparator Function : Ranking display of judgments 16 D. Comparator + Accumulation : Display of sum of judgment data...
SETTING OF JUDGMENT RANGE - In case Comparator Function is selected T ) After completion of selection of Comparator Function or Comparator + Accumulation Function by (¥) in page 13, press the [T] key to display El £"•.__,, which is for setting contents Comparator contents of the comparator function. 5 I. aC. I * For the details of the contents, see page 22. T) Jj Fisrt, set the judgment condition by pressing the Q J key, which changes the parameter by every pressing of it.
ACCUMULATION FUNCTION The Accumulation Function of the HG scale offers summing-up and display of measured data repeatedly, usuful for mixing or filling. OPERATION (1) Load an object on the scale. After stabilization mark ©appears, press the 0 0 key to accumulate the display value. The total amount will be seen for a moment with H mark in the display. Accumulation 100 pcs (B Total value 100 (2) Unload the object and allow the "0" display is settled down with © mark. Then load another object.
COMPARATOR FUNCTION SETTING WITH REFERENCE SAMPLE FOR JUDGMENT Setting reference points (1 point to 4 points) with samples, read ranking or bar graph for result of judgment. Beep is available. Convenient for measurement within a preset value. (1) Press the £s_J key until \_ SEb appears. M sign and some value will start blinking. Comparator function setting L. 5Et m The value shows that currently set.
MANUAL SETTING OF POINTS FOR JUDGMENT Setting.reference points (1 point to 4 points) by key operation, read ranking or bar graph for result of judgment. Beep is available. Convenient for measurement within a preset value. (1) (2) (3) Press the |_SJ key until L SEb appears. M sign and some value will start blinking. The value shows currently set value. While the display is blinking, press the | T | key to display "0" and bar graph underneath. Now the manual setting is starting.
RANKING INDICATION OF JUDGMENT RESULT 1. Display of reference value at each set point (Ex.) mmcsmmammcmmcaaacaa DDDGIi DDDCIf DDDDIt DDGGIt DDDDD ODDDD 1st point 2nd 3rd 4th point point point In the L 5E t mode, pressing the [jsj key will read currently set reference value at each point, which is shown by a bar at corresponding position as in the above illustration. 2.
VARIATIONS OF MEASUREMENT MODE in combination of functions You can select any scale mode from the following l i s t in combination of HG functions. 0 = effective FUNCTION MEASUREMENT MODE SCALE Remarks MODE unit ORDINARY g WEIGHING Z q PCS COUNTING z pcs MODE q pcs PERCENTAGE MODE 0 Counting 0 0 - 0 - 0 0 0 0 - 0 - A B 0 - A Accumulation of pcs.
CONTENTS OF VARIOUS FUNCTIONS PART I •fr mark shows the status set at the shipping time.
PART I - 2 VARIOUS WEIGHT UNITS & READABILITY SELECTION (Function 71 to 74) FUNCTIONS CONTENTS DISPLAY • / i 3 \ 3 WEIGHT UNITS GROUP "A" 1 1 uR * LEAST READABILITY mg g kg ct j5 oz IS / 1b FT ne Ex.HG-2000 3 HE 0.01g 0.02g dR 0.05g ..„. 0.1 g "s" a WEIGHT UNITS GROUP " B " ( Unit "B 1 13 ub is settable only n ORDINARY WEIGHING MODE.) * LEAST READABILITY 1H db Co a r s e 0.2 g No u n i t i mg 3 3 g kg H ct 5 oz 5 "Ib / c Fin e displayed Ex. HG-2000 °-01g 0.
CONTENTS OF VARIOUS FUNCTIONS PART II CONTENTS OF COMPARATOR FUNCTIONS (Function 21 to 25) In case the scale is set in the Comparator Mode 3 5 E L 3 or E S S L H following functions will be displayed before displaying 3 R Q <_, . FUNCTIONS JUDGEMENT CONTROL DISPLAY 3I r=* JUDGEMENT RANGE EE L ,r / E .un i i .1 SETTING POINTS DPJ C D n r E :3 .a a EH BEEP CONTROL *1 bu i 3 3 H 5 a JUDGEMENT INDICATION ES C C * 1 E CONTENTS Constant judgement, even data is unsettled.
CONTENTS OF VARIOUS FUNCTIONS PART III MEASUREMENT CONDITIONS & RESULTS, IN RELATION TO INTEGRATION TIME & STABILIZATION SPEED SETTING Integration Time Conditions Good S rE 5 Sd W 1 1 Normal S rE i B Sd 3 i No good Measurement Precise S rE H S L/ Normal • i Coarse S rB S Sd r E I B Sd 1 $ B Sd 3 3 i 5 3 I I i PART IV r E Stabilization Speed 1 B Sd H 1 Results Accurate i Not very accurate Slow I Quick CONTENTS OF INTERFACE CONTENTS No output 0 a : a c I Constant serial o
LIST OF VARIOUS MESSAGES DISPLAYS Appears by pressing [7] key : Func : To call functions, scale modes, calibration mode. U. SEt '• To start sampling (unit weight setting) in COUNTING mode. P- SEt • To start setting of reference value for 100% in PERCENTAGE mode. C. SEt : To start setting of constant in CONVERSION mode. CRL : To start calibration (span adjustment).* Appears by pressing | s | key : L 5Et : To start setting reference points in COMPARATOR mode.
TROUBLESHOOTINGS SYMPTOMS Setting of judgment points impossible Display does not chanae with a loa i Display is unstable Erroneous value reads in display Wrong linearity No display LOW-BATT lamp lights SLEEP lamp lights b-Err appears Unavailable weighing upto the capacity. a ~E r r appears u ~ E r~ r appears CAUSES & REMEDY * Function setting is erroneous. See page 13. * Reference value is over the capacity of the scale.
CALIBRATION (SPAN ADJUSTMENT) * When the function lock (switch inside) is effective, ie, in CE mode, Calibration is unavailable by the user. Contact the shipper. To achieve optimum accuracy from the scale, it should be calibrated in the area in which it is used, and recalibrated when it is relocated to other area. The following calibration procedure is simple, and not subject to operator errors, but does require a reference weight of the full capacity of the scale.
NOTES 1. To stop the calibration operation, press the [s_] key. 2. a ~Err signs that the reference weight is over the full capacity. 3. The calibration is available with a reference weight of over 50% of the scale capacity. Nevertheless, we recommend to adjust by a full capacity weight. I-Err will appear if the load is less than 50% of the capacity. 4. £ - £ r r w-j]i appear if the error exceeds 1% of the capacity, or any other object than a reference weight. Or, perhaps the scale is defective.
OUTPUT FORMAT VIBRA HG s c a l e provide a serial data transmission output. These signals are t r a n s m i t t e d by DIN type 8-pin connector on the rear panel of the s c a l e . Pin 1 Pin Pin Pin Pin 4 Transmitted Data Line 2 5 Signal Ground 3 Only Pin 4 and Pin 5 are used for the scale's data transmission as shown by above illustration. Other pins are unavailable, because they are occupied by the scale circuit.
ASCII CODES (EXTRACTION) Complete, listings of the ASCII codes are available, but the following list will be limited to those codes used in SHINKO scales. Octal Code 12 15 40 53 55 56 60 61 Octal Code Decimal Decimal Hexadecimal Character 50 LF Line Feed 62 OA 10 51 OD CR Carriaae Ret .