System Requirements
Gas Specications
For Model D-2
Detector Mode
PDECD PDHID Ar-PDPID Kr-PDPID Xe-PDPID
Discharge gas
Helium Helium 2% Ar in He 1.5% Kr in He 0.8% Xe in He
Carrier gas
Helium Helium * ** **
Dopant gas
3% Xe in He n/a n/a n/a n/a
Detector Mode
PDHID Ar-PDPID Kr-PDPID Xe-PDPID
Discharge gas
Helium 2% Ar in He 1.5% Kr in He 0.8% Xe in He
Carrier gas
Helium * ** **
** Any gas including He which has an ionization potential greater than 12 eV
** Any gas including He which has an ionization potential greater than 11 eV
The helium purifier (HP2) supplied with the system must be used for the discharge gas.
Purity Specifications
•Helium (discharge and carrier gas) must have a minimum purity of
99.999%, with < 20 ppm Ne impurity. For trace analysis of fixed gases,
we strongly recomment 99.9999% purity He with < 0.5 ppm Ne.
Ar-PDPID mode: 2% ± 0.3% Ar in 99.999% He balance
Kr-PDPID mode: 1.5% ± 0.1% Kr in 99.999% He balance
Xe-PDPID mode: 0.8% ± 0.2% Xe in 99.999% He balance
For Model D-2-I
** Any gas including He which has an ionization potential greater than 12 eV
** Any gas including He which has an ionization potential greater than 11 eV
Purity Specifications
•Helium (discharge and carrier gas) must have a minimum purity of
99.999%, with < 20 ppm Ne impurity. For trace analysis of fixed gases,
we strongly recomment 99.9999% purity He with < 0.5 ppm Ne.
•Ar-PDPID mode: 2% ± 0.2% Ar in 99.999% He balance
•Kr-PDPID mode: 1.5% ± 0.1% Kr in 99.999% He balance
•Xe-PDPID mode: 0.8% ± 0.2% Xe in 99.999% He balance