System information

Troubleshooting
Vicon MX Hardware System Reference F-3
Check the electrical continuity of the shielding between the analog
data generating device output and the Vicon patch panel with a
multimeter.
Test the Vicon capture system by injecting a known signal into it
via the patch panel (e.g. connect a 9V battery—one of the small
rectangular ones—or known voltage across an input channel and
ground).
Check the continuity of the connecting cables (100-way cable
between Vicon patch panel and MX Control, interface cables from
force plates and EMG equipment) with a multimeter to ensure no
breakages are present.
Use an oscilloscope to measure the signal present at the analog
data generating device output connector.
Ensure that all analog data generating devices (force plate
amplifier, EMG equipment, etc.) share a common signal ground
by correctly installing the interface cables such that the shields are
connected to a SGND screw terminal on the Vicon patch panel.