VIVIX-S 1012N User Manual
3
Close the Image dialog box, and open the Calibration dialog box.
4
Choose Defect Calibration as follows and click Manual Defect Correction button.
5
Move to the folder where the raw file is saved and select the file.
6
The Manual Defect Detection screen will be displayed.
7
Click Show and Apply buttons on the top left and then click Auto Defect Finder button.