User's Manual
Table Of Contents
- Instruction
- Product
- System Configuration
- Installation
- Setting
- Calibration
- Diagnosis, Inspection and Maintenance
- Troubleshooting
- Regulatory Information
- Medical Equipment Safety Standards
- Radio Frequency Compliance Information
- Labels and Symbols
- Guidance and Manufacturer’s Declaration for EMC
- Electromagnetic Emissions
- Electromagnetic Immunity
- Electrostatic Discharge (ESD) IEC 61000-4-2
- Electrical Fast Transient/Burst IEC 61000-4-4
- Surge IEC 61000-4-5
- Voltage Dips, Short Interruptions/Voltage Variations on Power Supply Input Lines IEC 61000-4-11
- Power Frequency (50/60 ㎐) Magnetic Field IEC 61000-4-8
- Conducted RF IEC 61000-4-6 / Radiated RF IEC 61000-4-3
- Information
![](/manual/vieworks/fxrd-1012naw/user-s-manual-english-romanian/images/img-115.png)
Rev.1.0
Page 115 of 151 VW40-152-009
VIVIX-S 1012N User Manual
7.1 Diagnosis
7.1.1 Image Diagnosis
Check the image quality through Diagnosis tools after installing the detector or before
usage. If the problems with regard to products or image occur during diagnosis, try to do
a calibration again. If the problems are not solved, consult the sales representative in
Vieworks or a service engineer.
You can acquire and review an image from the Diagnosis window in VIVIX Setup program.
You can review images by acquiring them through real exposure or getting Dark image by clicking Get
Normal Image button.
The number of images, pixel value and ROI value will be displayed.
The effective area or whole area of an image can be checked.
It is also available to check the image by changing its direction.
Save the reviewed image as a raw one to analyze.
You can acquire an image either applying Offset / Gain data or not.