Dignostic Functions
Item
Description
Pan
Press and drag a mouse button to m
ove the image to the desired position.
Zoom
Press and hold
a mouse button to move the image upwards or downwards for
expanding / reducing the image.
W/L
Press and hold
a mouse button to move the image to up/down/right/left for
adjusting its window level. This function can be used without clicking W/L button,
but with the right mouse button basically.
Fit
S
ets the image to the center, which was moved by using Pan function.
Statistic
Clicks and
drags the left mouse button to set arbitrary area. The coordinate,
min/max value, average and standard deviations are displayed on
Pickup ROI at
the left side of image.
ROI
Clicks and
drags the left mouse button to set window level automatically on a basis
of min/max value of the configured area
.
Zoom
(X1 / X2 / X4)
Expands the image (2 times / 4 times / 16 times)
Effective Area
Applicabilit
y of Effective Area in Detector Configuration dialog.
On
Applies effective area of the
detector to the image.
Off
Applies whole area of the
detector to the image.
Offset
Applicability of Offset data
On
Acquires an image with applying Offset data.
Off
Acquires an image without applying Offset data.
Gain
Applicability of Gain data
On
Acquires an image with applying Gain data.
Off
Acquires an image without
applying Gain data.
Direction
Applicability of
Detector Direction Compensation in Detector Configuration dialog.
On
The image is
displayed according to the configured direction.
Off
The image is displayed as a default direction. The image starting point (0,0) is
located on the left top of the image.
Get Normal Image
Acquires a d
ark image without X-ray shooting.
Save Image
Saves the image as a raw one.
Close
Close
s the Diagnosis dialog.
You can only change the displayed direction of an image in VIVIX Setup program from
Detector Direction Compensation. The direction of an original image cannot be
changed.