User's Manual

Table Of Contents
Rev.1.0
Page 132 of 151 VW40-152-009
VIVIX-S 1012N User Manual
7
Replace corresponding devices.
8.1.5 Errors in Detector LED
Symptom
All LED lamps of detector are blinking.
2 LED lamps of detector are blinking and remaining is blinking slowly.
Expected Causes
Internal hardware errors of the detector.
Solutions
1
Boot up the detector again and check the result of it
2
Replace the detector.
8.1.6 Rapid Consumption of Battery
Symptom
Consumption of fully charged battery pack is fast.
Expected Causes
Performance decrease caused by usage of long time.
Usage of battery pack in low temperature environment.
Solutions
1
Replace to new battery pack if the battery pack has been used for a long time. (Battery pack is
consumables)
2
Use battery pack in room temperature environment. Charging capacity of battery pack in low
temperature environment will be decreasing.
8.1.7 Battery Pack or Installation Part of Battery is Getting Hot
Symptom
Battery pack or compartment for installation of battery pack is getting hot.
Expected Causes
Battery pack failure
Detector Failure
Solutions
1
Do not use battery pack
2
Consult with service engineers of Vieworks.