Rev.1.0
Page 33 of 151 VW40-152-009
VIVIX-S 1012N User Manual
2.3.5 Wireless Communication
Item
Specifications
Wireless standard
IEEE802.11n
Frequency range
2.412 ~ 2.472
㎓ (13 Channels)
5.18 ~ 5.24
㎓ (4 Channels)
5.745 ~
5.805㎓ (4 Channels)
Data transmission rate
802.11n: Max. 300Mbps (MIMO 2X2)
Modulation
BPSK, QPSK, 16
-QAM, 64-QAM
Transmission power
Max. 17dBm
Security
WPA2
-PSK
Antenna
Dual Band Antennas (2EA, inbuilt)
The specification of detector and SCU are same, but the location of antenna is different.
2.3.6 Use Environment
Item
Operation
Storage
& Transportation
Temperature
+10 ~ +35℃
-
15 ~ +55℃
Humidity
30 ~ 85% (Non
-condensing)
10 ~ 90% (Non
-condensing)
Atmospheric pressure
700 ~ 1060 hPa
500 ~ 1060
hPa
Shock
1.6G
20G
Vibration
0.7G
0.7G
The use environment of detector and SCU is same.
Load Limit of Detector
Uniform load
Local load
Over the whole surface Center diameter 40mm
Max. 100 kg Max. 150 kg
Do not let the paitent or object heavier than load limit be on the detector. Then, detector
can be damaged.
Do not let the patient lie or get on the detector. Internal devices such as a sensor can be
seriously damaged even if his/her weight is within the load limit.