VIVIX-S 1012N User Manual
3.4.2 DR Trigger Interface
Connect the detector and X-ray generator with a generator interface cable, and then acquire images by
sending and receiving their signal.
Signal Processing Steps
1
The detector receives EXP_REQ signal from the X-ray generator.
2
After the detector completes to prepare image acquisition, it sends the EXP_OK signal to the X-ray
generator.
3
The X-ray generator confirms EXP_OK signal and generates X-rays.
4
The detector acquires images and then transmits the image data.
EXP_REQ is a sending signal from X-ray generator to the detector for requesting
exposure.
EXP_OK is an exposure ready signal from the detector to X-ray generator.