User's Manual

Table Of Contents
Rev.1.0
Page 67 of 151 VW40-152-009
VIVIX-S 1012N User Manual
4.2 Product Installation
4.2.1 Connecting Devices
Installation of this equipment should be made by licensed and authorized personnel by
Vieworks.
This equipment must only be connected to the power with protective earth.
SCU Basic (FXRS-02A)
1
If you use the DR Trigger mode, connect the one end of generator interface cable to the EXT_INF port
of SCU, and the other to the X-ray generator.
If you use the AED mode, a generator interface cable is not needed as the detector
operates by detecting X-ray automatically.
2
Connect one end of the LAN cable to one of the LAN ports of SCU, and the other to the LAN Card
Connector of workstation assigned for data transfer.
3
Connect the one end of the Tether Interface cable to the Port 4 or Port 5 of SCU and the other to the
detector.
4
To supply power, connect the AC power cable to the AC power input port of SCU.