User's Manual

Table Of Contents
Rev.1.0
Page 70 of 151 VW40-152-009
VIVIX-S 1012N User Manual
4.2.2 Booting up SCU
1
Turn on the power switch at the front side of SCU.
2
Check if the power switch of SCU lights up green.
3
Check if the status indicator is blinking in green after the power switch of SCU is permitted.
4
Check if the status indicator of SCU is blinking in green and turns to light up in the same color.
When the power switch lights up green, it means that the power is applied normally.
When the status indicator is blinking in green, the SCU is booting.
If the status indicator lights up green, it indicates that the SCU is being operated normally
after booting up.
If the status indicator lights up blue, SCUs preparation for communication is completed.
4.2.3 Booting up the Detector
The procedures in this chapter are explained according to the default setting status of a
detector (factory reset) and the use of wireless connection.
Refer to <7.4 Product Initialization> for the detailed information about the default setting
status of a detector.
1
Attach a charged battery pack to the detector.
2
Press the power button of the detector for 1 second until the power LED is turned on.
3
When the power LED is on, check if it is lit green.
4
Check if the active LED is lit orange on the detector.
If the power LED lights up green, the DC power is permitted normally.
If the active LED lights up orange, the detector is compeleted to boot up normally.