User's Manual

Table Of Contents
Rev.1.0
Page 88 of 151 VW40-152-009
VIVIX-S 1012N User Manual
WNetwork
Sets AP (Access Point) information of wireless communication from the detector.
Item
Description
SSID
Configures w
ireless network ID AP from the detector.
Key
Configures w
ireless network key value of AP from the detector.
Wireless Only
Configures the
wireless communication method of the detector.
On
The detector is operated in a wired way when a tether interface cable is
connected.
Off
The detector is operated in a wireless way when a tether interface cable is not
connected.
AP
Checks and sets AP (Access Point) of detector information.
Item
Description
AP On
/ Off
Decides w
hether to run the detector on AP (Access Point) mode.
On
Turns on
the detector AP
Off
Turns off
the detector AP
Frequency
F
requency channel of wireless network
2.4
Use
s 2.4 Frequency. (Up to 13 channels)
5
Uses 5
Frequency. (Up to 8 channels)
Country
C
ountry code of wireless network (KR, US, EU, JP, CN)
Band
W
ireless network bandwidth
20
Basic
Frequency Band
40
Expands
bandwidth through channel bonding.
Channel
W
ireless network channel
SSID
W
ireless network ID
Key
W
ireless network key value (Applied to the password only.)
13 channels can be used in 2.4 Frequency.
8 channels can be used in 5 Frequency
The number of serviceable channels is different according to the configured country.
Channel bonding is used for enhancing transmission speed. However, the speed may be
slowed down due to the interference of surrounding channels, even if the channels have
been bonded.
Channel items (+/-) will be activated in case of using 40 frequency bandwidth. You can
set whether to bond channels with the above or below one.
SSID and Key values of the detector for communicating with SCU wirelessly (WNetwork)
should be set as the same.