User's Manual

Table Of Contents
Rev.1.0
Page 94 of 151 VW40-152-009
VIVIX-S 1012N User Manual
6.1 Calibration Dialogue
Use the following menus to configure the system and process calibrations.
Menu
Description
System Configuration
Configures the exposure mode, exposure
-related time and Gain type.
Offset Calibration
Processes
Offset calibration.
Defect Calibration
Processes
calibration for defect data of the detector.
Gain Calibration
Processes
calibration for image sensitivity.
Detector Configuration
Configures the detector information related to calibration and image.
The menu status is displayed as Done when each calibration is finished. You cannot
operate the detector if the status is not displayed as Done.
6.1.1 System Configuration Dialogue
Configure the generator interface information from System Configuration in Step.
Exposure Mode
The VIVIX-S 1012N detector provides two types of mode for taking images as follows.
Mode
Description
AED
Detects X
-ray automatically without connecting the detector to X-ray generator.
The detector acquires images after the automatic detection.
DR Trigger
D
etects the X-ray exposure signal of X-ray generator in advance by connecting the
detector to X
-ray generator with a generator interface cable.
The
detector acquires images after preparing exposure with the exposure signal.