VIVIX-S 1012N User Manual
Exposure Timing
To take images exactly, set the timing information for exposure.
The information can be different depening on the characteristics of the X-ray generator.
Exposure Timing Diagram
Exposure Timing Items
Item
Default
Description
Exposure section
50
0
T
ime to acquire X-ray by the detector.
Pre Exposure section
0
Standby time after the detector sent EXP_OK signal to the generator.
EXP OK Delay section
1
The delayed time to receive EXP_OK signal from the detector after the
generator sent EXP_REQ signal to the detecto
r.
Debounce_DR
3
The required time to
check signals from the generator for preventing
image
acquisition from external noise. It can be used when DR Trigger
mode is selected only
.
Debounce_AED
3
The required time to check sensor signal from X
-ray to prevent image
acquisition
from external noise. It can be used when AED mode is
selected only
.
Exposure section should be set longer than the maximum exposure time to prevent X-
ray
loss.