User's Manual
Table Of Contents
- Instruction
- Product
- System Configuration
- Installation
- Setting
- Calibration
- Diagnosis, Inspection and Maintenance
- Troubleshooting
- Regulatory Information
- Medical Equipment Safety Standards
- Radio Frequency Compliance Information
- Labels and Symbols
- Guidance and Manufacturer’s Declaration for EMC
- Electromagnetic Emissions
- Electromagnetic Immunity
- Electrostatic Discharge (ESD) IEC 61000-4-2
- Electrical Fast Transient/Burst IEC 61000-4-4
- Surge IEC 61000-4-5
- Voltage Dips, Short Interruptions/Voltage Variations on Power Supply Input Lines IEC 61000-4-11
- Power Frequency (50/60 ㎐) Magnetic Field IEC 61000-4-8
- Conducted RF IEC 61000-4-6 / Radiated RF IEC 61000-4-3
- Information
![](/manual/vieworks/fxrd-1012naw/user-s-manual-english-romanian/images/img-95.png)
Rev.1.0
Page 95 of 151 VW40-152-009
VIVIX-S 1012N User Manual
Exposure Timing
To take images exactly, set the timing information for exposure.
The information can be different depening on the characteristics of the X-ray generator.
Exposure Timing Diagram
Exposure Timing Items
Item
Default
Description
Exposure section
50
0㎳
T
ime to acquire X-ray by the detector.
Pre Exposure section
0
㎳
Standby time after the detector sent EXP_OK signal to the generator.
EXP OK Delay section
1
㎳
The delayed time to receive EXP_OK signal from the detector after the
generator sent EXP_REQ signal to the detecto
r.
Debounce_DR
3
㎳
The required time to
check signals from the generator for preventing
image
acquisition from external noise. It can be used when DR Trigger
mode is selected only
.
Debounce_AED
3
㎳
The required time to check sensor signal from X
-ray to prevent image
acquisition
from external noise. It can be used when AED mode is
selected only
.
Exposure section should be set longer than the maximum exposure time to prevent X-
ray
loss.