6.1.4 Gain Calibration Dialogue
Vieworks provides the Gain data stored in the detector. However, you should download
the data as the calibration cannot be performed in the detector automatically.
This data is provided in case the Gain calibration cannot be performed. You should do
the Gain calibration in person as the detector condition can be different by the operating
method or use environment.
Gain Calibration
Item
Description
Target
Value
Shows the target value of Gain calibration. (Recommended :
8000 or higher)
Current Value
Shows
value after the exposure when processing the manual Gain calibration.
S
tage
Select
s the number of manual Gain calibration and shows each calibration stage.
Get
Starts the manual Gain calibration.
Cancel
Cancels
the manual Gain calibration.
Load Gain
Loads the Gain calibration data
located in the Gain path.
Upload Gain
Uploads the Gain calibration data to the detector for using portable mode.
Download Gain
Downloads the Gain calibration
data stored in the detector.
Gain
Path
Shows the
path of applied Gain data.