VIVIX-S 1417N User Manual
1.5.9 During Exposure
This equipment is not protected (sealed) against liquids such as blood and medicine in the operating
room. If necessary, wrap the equipment in a disposable cover when using it.
Do not use the selected frequency channel (2.4 and 5 dual band) for other wireless devices.
Mutual interference may affect the image data transmission rate.
Do not use the detector near devices generating a strong magnetic field. Doing so may produce image
noise or artifacts.
1.5.10 Operating and Storage Environment
This equipment is mainly used in the X-ray exposure room and hospital wards. To use it in other places,
consult Vieworks sales representative or a distributor.
Do not expose this equipment to high temperatures and/or high humidity. Malfunctions may occur.
When not in use, keep the products in a safe location.
Be sure to use and store this equipment under the conditions described below.
Classification
Humidity (Non-condensing)
Atmosphere
Operating Environment
30 ~ 85%
700 1060 hPa
Storage and Shipping
Environment
10 ~ 90%
500 1060 hPa
The operating environment of this equipment is complied with EN60721-3-3 Class 3K2.
The storage and shipping environment of this equipment is complied with EN60721-3-2
Class 2K2.
1.5.11 Others
Do not use this equipment in combination with peripherals such as defibrillators or large electric motors
as these may cause power-supply noise or power supply voltage variations. Doing so may prevent
normal operation of this equipment and peripherals.