User's Manual Part 1
Rev.1.5
Page 26 of 121 D-16-207
VIVIX-S 1417N User Manual
2.2 VIVIX-S 1417N Detector
VIVIX-S 1417N is designed to acquire digital images by collecting x-ray signals and sereval conversion
processes. You can use the acquired image (14” x 17” film size) diversely depending on the purpose of use.
2.2.1 Specifications
Item
Specifications
Model
FXRD-1417NAW (CsI)
FXRD-1417NBW (Gadox)
Image Sensor
TFT: a-Si (Amorphous Silicon)
X-ray Scintillator Type
FXRD-1417NAW : Csl: TI (Thallium doped Caesuim Iodide)
FXRD-1417NBW : Gd2O2S:Tb (Gadolinium oxysulfide)
Pixel Pitch
0.14㎜ (140㎛)
Field of View
14” x 17”
Active Area (H x V)
358.4㎜ × 430.08㎜
Active Array
2560 x 3072 pixels
Effective Area
FXRD-1417NAW: 355.04㎜ x 426.72㎜
FXRD-1417NBW: 358.4㎜ × 430.08㎜
Effective Array
FXRD-1417NAW: 2536 x 3048 pixels
FXRD-1417NBW: 2560 x 3072 pixels
Grayscale
16bit
Spatial Resolution
Min. 3.5 lp/㎜
Image Acquisition Time (Wired)
1.5 sec.
Image Acquisition Time (Wireless)
3 sec.
Recommended Cycle Time
10 sec.
X-ray Synchronization Control
AED (Auto Exposure Detection)
DR Trigger (External line trigger)
Rated Power Supply
DC +24V, Max. 1.0A
Wired: Powered by SCU with a tether interface cable.
Wireless: Powered by a battery pack (3,100 ㎃h x 2)
Power Consumption
Max. 19.2 W
Operating Time
6 hours (Capturing), 8 hours (Standby) - when sleep mode is off
Dimensions (H × W × D)
384.0㎜ × 460.0㎜ × 15.0㎜
Weight
3.0 kg (Without a battery pack)
Image Transfer
Wired: Gigabit Ethernet (1000BASE-T) via PoE (Power over
Ethernet)
Wireless: IEEE 802.11n/ac (2.4GHz/5GHz)
Data Transmission Rate (Wired)
Max. 1Gbps
Data Transmission Rate (Wireless)
IEEE802.11n: Max. 450Mbps (MIMO 3X3)