User's Manual Part 1

Rev.1.5
Page 66 of 121 D-16-207
VIVIX-S 1417N User Manual
Charging type 2
Before charging a battery pack, supply power to the charger with a power adapter.
1
Open the cover at the lower side of charger.
2
Insert the FXRB-03A battery pack into the slot until it clicks.
Insert the battery pack in the direction of a ‘TOP’ printed part.
3
Close the cover at the lower side of charger.
The charging time can be different depending on temperature and charging status of a
battery.
While charging a battery pack, it is available to use the detector by connecting it with a
tether interface cable.
This detector is designed to make a battery being discharged little by little during the
detector is off. Therefore, remove the battery pack when the detector is not used for
some time. Otherwise, over discharge may occur resulting in the shortened battery life.
Do not use the charger around the paitent.
Charge the battery with the charger provided by Vieworks.
Securely plug the power cord of the charger into the AC outlet. If contact failure occurs,
or if dust/metal objects come into contact with the exposed metal prongs of the plug, fire
or electric shock may result.
Use only a power adapter complying with IEC 60601-1 or IEC 60950-1.