VIVIX-S 1417N User Manual
4.2 Product Installation
This equipment should be installed by a service personnel authorized by Vieworks.
This equipment must only be connected to the power with protective earth.
4.2.1 Connecting SCU Basic (FXRS-03A)
1
If you use the DR Trigger interface, connect the one end of generator interface cable to the EXT_INF
port of SCU, and the other to the X-ray generator.
If you use the AED interface, a generator interface cable is not needed as the detector
operates by detecting X-ray automatically.
2
Connect one end of the LAN cable to one of the LAN ports of SCU, and the other to the LAN Card
Connector of workstation assigned for data transfer.
3
Stand the antenna of SCU upright.
4
To transmit image data using a tether interface, connect the one end of the Tether Interface cable to the
Port 4 or Port 5 of SCU and the other to the detector.