User's Manual Part 1
Rev.1.5
Page 72 of 121 D-16-207
VIVIX-S 1417N User Manual
4
To supply power, connect the DC power cable to the DC power input port of SCU mini.
4.2.3 Booting up SCU (Ex. FXRS-03A)
1
Turn on the power switch at the front side of SCU.
2
Check if the power switch of SCU lights up green.
3
Check if the status indicator is blinking in green after the power switch of SCU is permitted.
4
Check if the status indicator of SCU is blinking in green and turns to light up in the same color.
When the power switch lights up green, it means that the power is applied normally.
When the status indicator is blinking in green, the SCU is booting.
If the status indicator lights up green, it indicates that the SCU is being operated normally
after booting up.
4.2.4 Booting up the Detector
The procedures in this chapter are explained according to the default setting status of a
detector (factory reset) and the use of wireless connection.
Refer to <5.3.3 Wireless Initialization of Detector> for the detailed information about the
default setting status of a detector.
1
Attach a charged battery pack to the detector.
2
Press the power button of the detector for 1 second until the power LED is turned on.
3
When the power LED is on, check if it is lit green.
4
Check if the active LED is lit orange on the detector.