VIVIX-S 1417N User Manual
4.4.4 Wired/Wireless Communication Speed Diagnosis
1
Execute VIVIX Setup and go to the Information tab of the Diagnosis dialog.
2
Click on the Download button in Image Transmission Time and check image transmission speed of the
detector.
3
Click on the Start button in Throughput Measurement to check the data traffic per setting time.
Refer to VIVIX Setup Operation Manual for the detailed information about the
communication speed diagnosis.
Be sure to check the communication environment if there is any problem occurs in the
communication speed. Contact the person in charge of service if the problem is related to
the communication module of a detector and SCU.
4.4.5 Self-Diagnosis
1
Execute VIVIX Setup and go to the Self Diagnosis tab in the Diagnosis dialog.
2
Check the desired items to diagnose from the Test Case list.
3
Click on the Execute Test button to perform self-diagnosis.
4
Check the status and result of diagnosis for each item in the Result dialog.
Click Save to File button to save the result of a diagnosis as a file and contact the service
engineer if any problem is found.