5.2 Cleaning and Disinfection
After using the detector and peripheral equipments for examination, use germicidal disinfecting wipes or
cloth with mild diluted disinfectant detergent to clean surfaces of the product.
In case the surface or narrow space of equipment is contaminated by contact with blood
or other body fluids of a patient, make sure to clean and disinfect it to protect the
patients and users from infection.
5.2.1 Recommended Detergent Foam
Recommended disinfectant wipe
Super Sani-cloth Plus Wipes by PDI
Sani-cloth Active Wipes multi Surface (Alchol Free/Sans alcool) by PDI
Sani-cloth CHG 2% by PDI
Cavi Wipes by Kerr Total Care
Sporicidal Wipes by Clinell
Universal Wipes by Clinell
Recommended disinfectant product
Sulfa’safe by Anios
Storage temperature: 5°C ~ 35°C
5.2.2 How to Use Detergent Foam
1
Prepare the disinfectant detergent and a clean and dry non-woven cloth.
2
Use the spray bottle to spray detergent to the cloth and clean the equipment.
3
Clean residue on the equipment with its power off.
4
Conduct cleaning once a week or in case of contamination.
Do not re-use wipes.
Be careful to use disinfectant detergent which can cause irritation to eyes and skin.
Use in well-ventilated areas, and wear gloves at all times.
Do not clean the equipment with its power on.
Do not use abrasive brush and scraper to clean the product.
Do not make liquid soak when cleaning battery bay and a connector on side of products.
Other Disinfectant detergent compliant to conditions listed below may be used following
proper procedures according to its own manual.
European Biocidal Products designed for surface disinfection (Directive 98/8/EC)
Detergent with composition of Didecyldimethylammonium chloride, polyhexamethylene
biguamide hydrochloride.