VIVIX-S 1417N User Manual
5.6 Replacing the Fuse of SCU
The description in the section is applicable to the SCU Basic model only.
There are 2 fuses attached on the SCU Basic for the purpose of electrical accident precaution, in case of
over current from external power input. Stop using the SCU immediately when the fuse is blown.
Fuse Information
Item
Specifications
Model
Littelfuse® 218002 (2EA)
Type
Time Lag Cartridge Fuse
Amp Rating
2A
Voltage Rating
250V
Pull the plug out and turn all the devices off before changing the fuse.
First, resolve the cause why the fuse is blown. Replace the fuse to the one provided as an
option (1 set / 2 ea) or to the one with same specifications when the fuse is out.
Be careful not to touch both the patient and the fuse holder at the same time or let the
patient touch the fuse holder.
This product has lower breaking capacity type. So do not install the product at the
building power system which prospective short-circuit current is exceeding 35 A.
How to Replace the Fuse
No.
Description
1
Separate the fuse from the holder located power input port on
the back side of standard SCU by pulling the fuse holder.
2
After checking, replace the fuse with correct specifications in
case of need.
3
-
Insert the fuse holder again.