VIVIX-S 1417N User Manual
6.1.4 The Power Switch of SCU Lite is not Working
Category
Description
Symptom
The LINK LED of SCU Lite is not working even though its power switch is turned
on.
Expected Causes
The power cable is broken down.
The internal circuit is broken down.
Solutions
1 Check the connection between DC power cable and SCU Lite.
2 Turn off the power switch and turn it on again. Check if the LINK LED of SCU Lite
is working properly.
3 Replace SCU Lite to another one and check the result.
6.1.5 Communication Test is failed
Category
Description
Symptom
Transmission error occurred, and the communication test is failed.
Expected Causes
Network connection problem
Network setting problem
PC environment setting problem
Wireless environment environment problem
Errors in devices
Solutions
1 Check the connection of network cable between Workstation and SCU.
2 Check if the accurate network cable is used or not. (CAT 5E or 6)
3 Set the network information of Workstation, SCU and detector again.
4 Check whole workstation environment again such as firewall setting and release
the power save mode.
5 Check surrounding wireless communication environment.
6 Reboot the detector and SCU again by processing initialization. (Refer to
<5.3Product Initialization>)
7 Replace the detector and SCU Basic to another ones and check the result of it.
8 Replace corresponding devices.