User's Manual Part 2

Rev.1.5
Page 98 of 121 D-16-207
VIVIX-S 1417N User Manual
6.1.6 The Active LED and Data LED of the Detector are blinking
Category
Description
Symptom
The active LED and data LED are blinking while power LED is turned ON.
Expected Causes
Detector registeration error
Data transmission error
Solutions
1 Turn on SCU again.
2 Check the network cable connection.
3 Check the workstation environment and network information again.
4 Check if the surrounding wireless communication is good.
5 Check cable connection again when connected with tether interface cabel
6 Replace other devices and check the result of it
7 Replace corresponding devices.
6.1.7 Errors in Detector LED
Category
Description
Symptom
All LEDs of a detector are blinking.
Two LED lamps of a detector are blinking and the remaining one is blinking slowly.
Expected Causes
Internal hardware errors of a detector.
Solutions
1 Reboot the detector and check the result.
2 Replace the detector to another one.
6.1.8 Rapid Consumption of Battery
Category
Description
Symptom
The fully-charged battery is consumed rapidly.
Expected Causes
The performance of a detector is decreased by its length of use.
In case of using a battery under in low temperature.
Solutions
1 Replace a battery pack if it is used for a long time. (Battery pack is consumables.)
2 Use a battery pack at room temperature. Charging capacity of a battery pack is
decreased in low temperature.
6.1.9 Battery Pack or Installation Part of Battery is Getting Hot
Category
Description
Symptom
A compartment for installing a battery pack is getting hot.
Expected Causes
A battery malfunction
Detector failure
Solutions
1 Stop using a battery pack.
2 Contact the service engineer in Vieworks.