VIVIX-S 1417N User Manual
4.4 Diagnosis of Devices
Execute the VIVIX Setup program to check if threre is any problem to operate the detector / SCU after
installing and setting devices.
Diagnosis Items
Items
Description
Image
Diagnoses the acquired images.
Battery Pack
Diagnoses the condition of a battery pack.
Wireless Communication
Diagnoses the status of wireless communication.
Wired/Wireless Communication Speed
Diagnoses the speed of wired/wireless communication.
Self-diagnosis
Diagnoses defects of a detector by self-diagnosis.
Set the devices and perform calibration again if any problem is found during the
diagnosis. Contact the person in charge of service if the problem is not corrected.
4.4.1 Image Diagnosis
1
Execute VIVIX Setup and move to the Image dialog.
2
Take an image and check if it has any problem.
3
Take a dark image and check if it has any problem.
4
Check the effective area and whole area of the image.
Refer to VIVIX Setup Operation Manual for the detailed information about the image
diagnosis.
If any problem is found on the image, check if it is caused by the surrounding
environment and calibrate the detector again. Contact the person in charge of Vieworks if
the problem is caused by the performance of a detector.