User's Manual

Rev.1.0
3.4.2
DR Trigger Interface Mode
Connect the detector and X-
ray generator with a generator interface cable, and then
sending and receiving their
signal
Signal Processing Steps
1
The detector receives
EXP_REQ
2
After the detector
complete
generator.
3
The X-
ray generator confirms
4
The detector acquires
image
EXP_REQ
is a sending signal from X
exposure.
EXP_OK
is an exposure ready signal from the detector to X
Page 49 of 109
VI
VIX
DR Trigger Interface Mode
ray generator with a generator interface cable, and then
signal
.
EXP_REQ
signal from the X-ray generator.
complete
s to prepare image acquisition, it sends the
EXP_OK
ray generator confirms
EXP_OK signal and generates X-rays.
image
s and then transmits the image data.
is a sending signal from X
-ray generator to the
detector for requesting
is an exposure ready signal from the detector to X
-
ray generator.
RA14-14B-011
VIX
-S 1012N User Manual
ray generator with a generator interface cable, and then
acquire images by
EXP_OK
signal to the X-ray
detector for requesting
ray generator.