VIVIX-S 1012N User Manual
24 EXT_W- Input Contact - - Returns signal from EXT_W+
25 EXT_X+ Input Contact - - User-defined Pin 2
26 EXT_X- Input Contact - - Returns signal from EXT_X+
27 EXT_Y+ Input Contact - - User-defined Pin 3
28 EXT_Y- Input Contact - - Returns signal from EXT_Y
29 EXT_Z+ Input Contact - - User-defined Pin 4
30 EXT_Z- Input Contact - - Returns signal from EXT_Z+
Since VIVIX SCU mini has EXT_INF 1 port, only one generator can be connected with.
Up to three X-ray generators can be connected to EXT_INF1 and EXT_INF2 ports, and
each generator should be connected with a same channel.
There are Contact type (OPEN/ CLOSE) and TTL type (ON/ OFF) on the signal in/output
pin.
TTL type information
ON : VCC
OFF : GND
Current: 5 ~ 10
Voltage: 12 V ~ 24 V
When you plan to add interface using a user-defined pin, contact to the person in charge
of Vieworks.