Check the connection between AC power cable and basic SCU.
2
Check the connection between DC power cable and mini SCU.
3
Turn off the power switch and turn on again and then check the fan or status of back side.
4
Replace another SCU and check the result of it
5
Replace the fuse of standard SCU. (refer to Change fuse of SCU)
6
Replace corresponding devices.
8.1.3 Communication Test is failed
Symptom
Transmission error is occurred, failure of communication test is occurred.
Expected Causes
Network connection problem
Network setting problem
PC environment setting problem
Wireless environment environment problem
Devices failure
Solutions
1
Check the connection of network cable between Workstation and SCU.
2
Check if the accurate network cable is used or not. (CAT 5E or 6)
3
Set the network information of Workstation, SCU and detector again.
4
Check whole workstation environment again such as firewall setting and release the power save mode.
5
Check surrounding wireless communication environment.
6
Boot up detector and SCU again by processing initialization. (Refer to <오류! 참조 원본을 찾을
없습니다. 오류! 참조 원본을 찾을 없습니다.>.)
7
Replace other SCU and Detectors and check the result of it
8
Replace corresponding devices.
8.1.4 The Active LED and Data LED of the Detector are blinking
Symtom
The active LED and data LED is blinking when power LED is ON.
Expected Causes
Detector registeration error
Data transmission error
Solutions
1
Turn on SCU again
2
Check the network cable connection
3
Check the workstation environment and network information again