User's Manual
  Wireless Service Manual 
  Page 100 of 124  RA14-11A-022 
5.4.4  Gain Pixel Correction 
Before performing Gain Pixel Correction, consider followings: 
  Recommended SID is 150  ㎝  (distance between X-ray tube and Detector). 
  Open the collimator of X-ray tube completely. 
  Align the center of the detector with the center of collimator. 
  Keep everything away from the detector surface. 
1  Select Gain acquisition – Normal in the left Step item. Expose X-rays and adjust dose of X-rays so 
that Current Value reaches TargetValue. Keep the quantity of X-rays when Current Value remains within 
10% difference of TargetValue. 
Recommended TargetValue is 2000. 
2  Click the Get button and expose X-rays ten times. 
3  Save the Gain data to the desired folder. 
4  Once all the procedures are completed, the OK button will be activated. Click the OK button to close the 
dialog box. 