User's Manual
  Wireless Service Manual 
  Page 108 of 124  RA14-11A-022 
Test ITEM (2) 
Temperature 
Frequency 
D: Daily  ■  M: Monthly  □ 
Q: Quarterly  □  A: Annually  □ 
Objective: 
Check the monitoring in order to minimize the characteristic changes of Flat panel caused by external 
temperature changes. 
Equipment: 
Workstation, VXvue S/W 
Power supply, Power Meter 
Inspection Report Form 
Procedure: 
Turn on the power of the system. 
Run VXvue. 
Check the temperature. 
Performance and Corrective Action: 
Confirm that the detector and ambient temperature is lower or higher than the operating temperature stated 
in this Service Manual. If the detector and ambient temperature deviates from the operating range, adjust 
the detector and ambient temperature properly to prevent poor image quality. If the image quality is reduced 
after adjusting the temperature, service assistance is necessary. 
Record result on Inspection Report Form. 
Result: 
Frequency: D: □,    M: □,    Q: □,    A: □  Date: 20**-**-**  Operator: 
Limit of Acceptability: +10 ~ +35℃ 
Remarks: 