User's Manual
  Wireless Service Manual 
  Page 110 of 124  RA14-11A-022 
Test ITEM (4) 
Work list connection 
Frequency 
D: Daily  ■  M: Monthly  □ 
Q: Quarterly  □  A: Annually  □ 
Objective: 
Confirm that VXvue is connected normally with Worklist and can register study by querying from Worklist 
server. 
Equipment: 
Workstation, VXvue S/W, Work List Server 
Power supply, Power Meter 
Inspection Report Form 
Procedure: 
Turn on the power of the system. 
Run VXvue. 
Select Setting Mode. 
Select DICOM tab. 
Select MWL tab. 
Select Worklist server. 
Click the Echo button. 
Performance and Corrective Action: 
Check the connection status that is displayed in test result window. 
If connection test failed, check Worklist server is operating and configuration of Worklist server connection 
is correct. 
If still Worklist connection fails, service assistance is necessary. 
Record result on Inspection Report Form. 
Result: 
Frequency: D: □,    M: □,    Q: □,    A: □  Date: 20**-**-**  Operator: 
Limit of Acceptability: N/A 
Remarks: 