User's Manual
  Wireless Service Manual 
  Page 111 of 124  RA14-11A-022 
Test ITEM (5) 
PACS Server connection 
Frequency 
D: Daily  ■  M: Monthly  □ 
Q: Quarterly  □  A: Annually  □ 
Objective: 
Confirm that VXvue is connected normally with PACS Server and can send performed study data to PACS 
server. 
Equipment: 
Workstation, VXvue S/W, PACS Server 
Power supply, Power Meter 
Inspection Report Form 
Procedure: 
Turn on the power of the system. 
Run VXvue. 
Select Setting Mode. 
Select DICOM tab. 
Select Storage tab. 
Select Storage Server. 
Click the Echo button. 
Performance and Corrective Action: 
Check the connection status that is displayed in test result window. 
If connection test failed, check PACS server is operating and configuration of PACS server connection is 
correct. 
If still PACS server connection fails, service assistance is necessary. 
Record result on Inspection Report Form. 
Result: 
Frequency: D: □,    M: □,    Q: □,    A: □  Date: 20**-**-**  Operator: 
Limit of Acceptability: N/A 
Remarks: 










